TOTAL ALKALINITY
REMOVE STRIP
Remove one eXact
®
Strip Micro AL Total Alkalinity
and set in a dry, convenient place. Replace cap
on bottle.
5
READ RESULTS
READ
result displayed as Total Alkalinity. This result is
automatically stored in the AL1 menu. After testing is
complete, rinse the sample cell immediately.
7
FILL, DIP, READ
DIP STRIP AND PRESS READ
Press
to initiate a 20 second countdown and
simultaneously
DIP
the eXact
®
Strip by submerging all
pads in the sample then use a gentle, constant, back
and forth motion (2 strokes per second) until the timer
displays
‘1’
. Remove and discard the strip.
6
If result reads "HI", repeat steps 5-7 with a Total Alkalinity Range
Extender Strip (part no. 486665) until a numerical value is
displayed; each strip adds 130 ppm to the
fi
nal result.
TIP
For water temperatures above 95°F/35°C (hot tubs), remove
the strip when the timer displays “10”, or use the specially
calibrated SPA eXact
®
EZ photometer.
TIP