5. Maintenance
DEPOLOX® Pool E flow cell
34
EN
11 Check that the check valve housing is cor-
rectly positioned by the guide lugs on the
housing.
12 Fit the filter unit again. To do this, tighten
the knurled nuts.
13 Refit the housing cover.
14 Connect the sample water inlet and outlet
again.
15 Reconnect the power supply.
Fig. 2
Section, installation position of the
check valve housing
Fig. 3
Cross-section of the check valve hou-
sing
A Check valve housing (overall view)
B
Flow ball
C
Ball seat
D Glass ball
5.7
Cleaning or replacing the
optional strainer
The optional strainer must be cleaned or
replaced regularly to avoid contamination or
blockages. The frequency of cleaning or repla-
cement depends on the degree of contamina-
tion caused by the sample water.
1 Close the shut-off ball valve on the sample
water inlet in front of the strainer.
2 Unscrew the strainer and rinse it with
water. Catch escaping water in a contai-
ner.
3 Remove the strainer screen and rinse it
under running water or replace it.
4 Fit the strainer screen again and reinstall
the strainer.
5 Open the shut-off ball valve on the sam-
ple water inlet and outlet again.
5.8
Cleaning
Never use corrosive cleaning agents (e.g. spi-
rit, scouring agents)! We recommend that you
use a moist cloth with a neutral household
cleaning agent.
A
top
bottom
B
C
D