DEPOLOX® Pool E flow cell
3. Description
EN
3.9.4 Sensors
NOTICE
When disinfecting inline electrolysis systems, the gold version of the chlorine sensor or the
ORP (Redox) single-rod measuring chain must be used.
Chlorine sensor
Platinum version (W3T160652)
Gold version (W3T160991)
Version
Amperometric 3-electrode
sensor with platinum
electrodes, salt reserve,
zirconium dioxide diaphragm,
polymerized solid electrolyte,
Ag/AgCl drain system
Amperometric 3-electrode
sensor with gold electrodes,
salt reserve, zirconium dioxide
diaphragm, polymerized solid
electrolyte, Ag/AgCl drain
system
Measurement range
0 to 50 mg/l Cl
2
Working temperature
range
-5 to +80°C (23 to 176°F)
Operating pressure
0 to 6 bar (6 x 10
5
Pa)
Minimum conductivity
of the sample water
50 µS/cm
Installation length
165 mm
Screw-in thread
PG 13.5
Storage temperature
-5 to +30°C (23 to 86°F)
ORP (Redox) single-rod
measuring chain
Platinum version (W3T169298)
Gold version (W3T172356)
Version
Single-rod measuring chain
with platinum electrode, salt
reserve, zirconium dioxide
diaphragm, polymerized solid
electrolyte, Ag/AgCl drain
system
Single-rod measuring chain
with gold electrode, salt
reserve, zirconium dioxide
diaphragm, polymerized solid
electrolyte, Ag/AgCl drain
system
Measurement range
±2000 mV
Working temperature
range
-5 to +80°C (23 to 176°F)
Operating pressure
(electrode shaft)
0 to 6 bar (6 x 10
5
Pa)
Minimum conductivity
of the sample water
50 µS/cm
Installation length
120 mm
Screw-in thread
PG 13.5
Storage temperature
-5 to +30°C (23 to 86°F)