G2 DISPLAY BOARD – IP-POWERDSP-G2
IP-POWERDSP-G2MAN Rev 1
Page 14 of 25
Figure 9: OVERVIEW Screen Example
RUN
: this submenu provides operational data and access to adjust the DC output current of
the selected power controller, as well as access to submenus SETUP, ALARMS, LOGS and
DIAGNOSTICS.
There are 5 pages in this submenu. Data is grouped as follows:
PAGE 1:
Figure 10: RUN Screen – Page 1
V DC out
: actual DC output voltage from the power controller to the selected
CEDI module.
A DC out
: setpoint for the DC output current from the power controller to the
selected CEDI module. Press the ENTER key to adjust this value using the UP
or DOWN keys. Press ENTER again once the desired value has been set. This
setpoint can not be set to a value outside the current range selected in the power
controller via DIP switch SW1.
Ohms pres
: actual resistance in ohms of the selected CEDI module. Resistance
values above 999.9 are represented with “***HI”. When the DC output current is
zero, resistance will be represented with “----“ to indicate that resistance can not
be measured.
Ohms init
: initial resistance in ohms of the selected CEDI module. This value is
captured when the power controller has operated continuously for 2 hours after