IONPURE 

®

 

G2 DISPLAY BOARD – IP-POWERDSP-G2

 

 

IP-POWERDSP-G2MAN Rev 1   

Page 14 of 25 

 

Figure 9: OVERVIEW Screen Example 

 

RUN

: this submenu provides operational data and access to adjust the DC output current of 

the selected power controller, as well as access to submenus SETUP, ALARMS, LOGS and 
DIAGNOSTICS. 

There are 5 pages in this submenu. Data is grouped as follows: 

 

PAGE 1: 

 

Figure 10: RUN Screen – Page 1 

 

V DC out

: actual DC output voltage from the power controller to the selected 

CEDI module. 

A DC out

: setpoint for the DC output current from the power controller to the 

selected CEDI module. Press the ENTER key to adjust this value using the UP 
or DOWN keys. Press ENTER again once the desired value has been set. This 
setpoint can not be set to a value outside the current range selected in the power 
controller via DIP switch SW1. 

Ohms pres

: actual resistance in ohms of the selected CEDI module. Resistance 

values above 999.9 are represented with “***HI”. When the DC output current is 
zero, resistance will be represented with “----“ to indicate that resistance can not 
be measured.  

Ohms init

: initial resistance in ohms of the selected CEDI module. This value is 

captured when the power controller has operated continuously for 2 hours after 

Содержание IONPURE IP-POWERDSP-G2

Страница 1: ...achusetts 01851 Tel 866 876 3340 Fax 978 934 9499 www ionpure com IONPURE G2 DISPLAY BOARD For use with G2 Power Controllers Operation Maintenance Manual IP POWERDSP G2MAN November 2011 Revision 1 Man...

Страница 2: ...2 3 2 Connections to Multiple G2 Power Controllers 9 2 4 Power Controller Setup 10 2 4 1 DC Output Current Range Selection 10 2 4 2 Unit ID Selection 11 3 OPERATION 12 3 1 Control Navigation Keys 12...

Страница 3: ...Screen Page 2 15 Figure 12 RUN Screen Page 3 15 Figure 13 RUN Screen Page 4 16 Figure 14 RUN Screen Page 5 16 Figure 15 SETUP Screen Page 1 17 Figure 16 SETUP Screen Page 2 18 Figure 17 ALARMS Screen...

Страница 4: ...ns has proprietary rights Neither receipt nor possession of this manual confers or transfers any right to the client and by its retention hereof the client acknowledges that it will not reproduce or c...

Страница 5: ...oid personal injury or fatalities Cautions indicate a situation that may cause damage or destruction of equipment or may pose a long term health hazard Notes are also used to draw attention to informa...

Страница 6: ...parameters have been configured the setup menu may be locked out to require a four digit password for further modifications One of the two RJ 45 ports available on the back of the unit is used for co...

Страница 7: ...rovide a seal at the panel cutout and it is secured from the back with the provided nuts The environmental limits for operation are 0 to 50 C at up to 95 relative humidity non condensing Consider the...

Страница 8: ...of burrs Figure 3 Panel Cutout Dimensions b Install the display board on the door of the enclosure using the supplied hardware and gasket 2 3 Connections 2 3 1 Connections to a Single G2 Power Contro...

Страница 9: ...s to Multiple G2 Power Controllers To connect the G2 display to multiple G2 power controllers a Connect one end of a standard Cat 6 cable to one of the two RJ45 ports on the back of the display b Conn...

Страница 10: ...ing excessive current to a CEDI module it is recommended to select an appropriate output current range with DIP switch SW1 of the G2 Power Controller as described in Figure 6 Select the current range...

Страница 11: ...hat is connected to a G2 display so they can be recognized without conflicts This is done via rotary switch SW2 of the G2 Power Controller Each SW2 position corresponds to a channel or module MD ident...

Страница 12: ...een submenus or configurable setpoints When adjusting a setpoint press to increase the value Hold to increase the value at a higher speed DOWN press to see the previous page of information or select b...

Страница 13: ...ine of asterisks Resistance values above 999 9 are represented with HI When the DC output current is zero resistance will be represented with to indicate that resistance can not be measured Channels 1...

Страница 14: ...C output current from the power controller to the selected CEDI module Press the ENTER key to adjust this value using the UP or DOWN keys Press ENTER again once the desired value has been set This set...

Страница 15: ...t to present Ohms pres Positive values represent an increase in resistance and negative values represent a decrease in resistance in reference to the initial value V AC in actual AC input voltage to t...

Страница 16: ...tween terminals Vm Vm of the selected power controller Im VDC estimated value of analog signal between terminals Im Im of the selected power controller PAGE 4 Figure 13 RUN Screen Page 4 Board temp te...

Страница 17: ...high resistance limit as follows once the module has come to steady state typically after about 48 hours of actual operating time set the Ohms limit to 1 5 times the steady state electrical resistance...

Страница 18: ...commended to use this option to reset all historic data after a module replacement Pressing ENTER will reset the following Op hours value in RUN submenu Ohms init value in RUN submenu Resistance alarm...

Страница 19: ...Resistance reports the level status of the actual resistance of the CEDI module The alarm will be ok if the resistance is lower than the value set for Ohms limit in the SETUP submenu it will be hi if...

Страница 20: ...LUES Screen VAC maximum registered AC input voltage to the selected power controller AAC maximum registered AC current drawn by the selected power controller VDC maximum registered DC output volts fro...

Страница 21: ...er VDC minimum registered DC output volts from the selected power controller to the CEDI module ADC minimum registered DC amps from the selected power controller to the CEDI module Ohms minimum regist...

Страница 22: ...f the power controller Display rev firmware revision of the display board Self test a G2 power controller monitors continuously the status of its power stage and temperature sensor The G2 display boar...

Страница 23: ...the dilute and concentrate streams through the module 4 For each operating channel on the G2 display enter the RUN submenu and adjust A DC out to set the DC output current to the value calculated by t...

Страница 24: ...ler have incompatible firmware revisions Contact parts supplier to update firmware A channel row is blinking Alarm has triggered Access ALARMS submenu to identify the alarm Two or more power controlle...

Страница 25: ...ents for Electrical Equipment for Measurement Control and Laboratory Use Part 1 General Requirements EUROPEAN UNION DIRECTIVE The G2 display board was evaluated as an accessory of the G2 power control...