Volatile organic compounds transmitter E2648-PID with photoionization detector is
intended for detection of organic vapors in very low concentrations.
The  device  belongs  to  the  PluraSens®  family  of  multifunctional  measurement
instruments.
E2648  series  provides  two  independent  analog  outputs  OUT1  and  OUT2,  user-
selectable to 4-20 mA or 0-10 V, proportional either to gas concentration or temperature.
RS485  Modbus  RTU  digital  communication  interface  allows  easy  instrument
configuration and integration into various automation systems.
The following features are available as option:

Two relays RE1 and RE2 with closing contacts. Relay outputs can be used to switch
alarm sirens, ventilation fans, shut-off valves or other actuators
Integrated 230 V mains power supply module
Remote probe.

The version of your device is marked on the package.
If  symbol

  is  marked  on  the  equipment,  consult  the  documentation  for  further

information.
PID technology
The gas under investigation diffuses into the sensor’s testing cell, equipped wit a
source of high energy UV light (10.6 eV Krypton lamp) and electrodes. Exposed to
UV  light,  organic  molecules  decompose  to  form  ions,  which  are  attracted  by  the
electrodes.  The  measured  resulting  current  is  proportional  to  the  VOC
concentration.
Safety requirements
Misuse will impair the protection of the equipment.

Always adhere to the safety provisions applicable in the country of use.
Do not perform any maintenance operation with the power on. Do not let water or
foreign objects inside the device.
External circuits connected to the equipment should have sufficient insulation rating
according to the environmental conditions and equipment power
A  disconnecting  device  that  is  marked  as  such  and  easily  accessible  should  be
included in the installation of this product.

Operating conditions
The device should be used in explosion-safe areas at the  temperature in the range
-40ºC  to  +65ºC,  15  to  95%  RH  without  condensation  and  under  ambient  pressure
0,9...1,1 atm.  Avoid  strong  mechanical  shock,  vibrations  or  EMI.  Avoid  corrosive
atmosphere and areas highly contaminated with dust, oil mist etc. Keep the device away
from direct sunlight.
Mounting dimensions

Installation and connections

Before  proceeding  with  the  installation  it  is  mandatory  to  read  carefully  the  Safety
requirements
 section and make sure to comply with all listed instructions. During the
installation of the detector the following points must be considered:
application (air quality control or leakage detection);
properties of the space under investigation (room geometry, direction and velocity of
air flows etc);
detected gas (relative density to air, temperature, whether the gas is flammable, or
toxiс, or oxygen displacing, whether the gas is heated);
the device should be accessible for maintenance and repair.

Solvent vapours are heavier than air and tend to sink. Do not take the gas samples for
investigation from too high.
The device is fixed on the wall using two holes located outside the sealed area of the
device(see dimensional drawing in the previous section).
Connections
1. Unscrew four lid screws and detach the lid from the device.
2. Attach the device to the wall. (This step may be done after the step 3, consider
your convenience).

3. Use two M16 cable glands to let in the cables of the power supply and of the external
devices.  Without  turning  on  the  power,  plug  the  power  cable  and  connect  the
analog/relay  outputs  and/or  digital  interface  terminals  to  the  necessary  devices
according to the relevant connection diagram.
The  screwless  quick  connect  spring  terminals  on  the  E2648  series  devices  are
suitable for a wide range of wires with cross-section 0,2...1,5 mm

2

. We recommend

to strip the wire end by 8...9 mm and use the wire end sleeves.

To connect the wire, insert the wire end into terminal hole. To disconnect, push the
spring loaded terminal lever, pull the wire out, and release the lever.
Use twisted pair cable, e.g. LiYY TP 2×2×0,5 mm

2

 or CAT 5, to connect the device

to  RS485  network.  Use  one  pair  for  A  and  B  wires  to  connect  the  transmitter  to
Fieldbus network. Respect polarity.
The type of each analog output can be independently changed between 4-20 mA
and 0-10 V with jumpers J1 (OUT1) and J2 (OUT2). With closed jumper the output
is 0-10 V, with open jumper the output is 4-20 mA. Power restart is required after
changing the position of the jumpers.
By default both outputs OUT1 and OUT2 are assigned to gas concentration. The
device has built-in temperature sensor which may be tied to any of the outputs. The
built-in temperature sensor is intended for temperature compensation. E2648 series
devices are not meant for ambient air temperature measurement.
We recommend to set the difference between the upper and bottom limits of the
output scale not narrower than 20% of detection range (for CO detectors the scales
down to 5% of range are allowed). In any case, do not set the output scale below
the tenfold resolution of the device. 
NOTE  The  outputs  are  not  galvanically  isolated  from  24  V  power  supply  and  share
common 0V. Allowed load resistance limits are stated in Specifications table. To power
the instrument from an external 24 VDC source, connect terminals 0V and +U to the
source. If the integrated mains power supply module is used, connect terminals L and N
to the mains.
NOTE  Actuator  short-circuits  should  be  avoided,  to  protect  the  instrument  relays
use external fuses or safety switches.
4. Turn on the power. The sensor heating up takes ca. 60 seconds after switching on.
The  operating  status  is  indicated  by  the  LED  on  the  PCB  of  the  device.  The  LED
response to different processes is presented in the table below.

Process

LED mode

Sensor absence or malfunction

Blinking 0.5 Hz (90% off, 10% on)

Relay1 turned on

Blinking 1 Hz (50% on, 50% off)

Relay 2 turned on

Blinking 2 Hz (50% on, 50% off)

Modbus response

The signal is modulated with short on-off pulses,
even single Modbus cycle is traceable*

Normal operating

Continuous light

5. Make sure that the device is properly installed, the external devices connected,
power on and control LED is constantly lit. Make certain that the cable glands are
properly tightened to ensure the conformity to IP66 protection class. Place the lid
back and fix it with the screws. The device is ready to use.
Configuring
Detectors- transmitters E2648 share all functionalities of the PluraSens

®

 multifunctional

platform. The features and options include:
- digital output change rate limiting filter
- digital integrating (averaging) filter
- temperature measurement channel with internal sensor
- free assignment of each analog output to chosen parameter
- flexible setting of analog output scales for each output
- output shift and slope adjustment for calibration
- free assignment of each of two relays to chosen parameter
- several relay control logic modes
- switch delays and minimum on/off state durations for each relay
A  standard  configuration  kit  includes  a  USB-RS485  converter  and  a  software  pack.
Please contact your Seller for more information.

E2648 version without relays

E2648R version with relays

Содержание PluraSens E2648-PID

Страница 1: ...erial and workmanship for a period of one year from the date of original sale During this warranty period Manufacturer will at its option either repair or replace product that proves to be defective T...

Страница 2: ...and or digital interface terminals to the necessary devices according to the relevant connection diagram The screwless quick connect spring terminals on the E2648 series devices are suitable for a wi...

Страница 3: ...may be addressed either from 00001 or 40001 base Addr Reg MHR RW Description Supported values dec Default 0x0001 1 40002 R Hardware version 0x0002 2 40003 R Software version 0x0003 3 40004 R Product s...

Страница 4: ...in MHR 40205 2 0x00CB 203 40204 RW Forced value for analog output 1 0 1000 0 0 100 0 of output scale 0 0x00CC 204 40205 RW Forced value for analog output 2 0 1000 0 0 100 0 of output scale 0 0x00D3 2...

Страница 5: ...0 1 LED is on off in normal condition bit 11 0 1 LED and buzzer syncing off on with RE1 bit 12 0 1 LED and buzzer syncing off on with RE2 user defined 0x0101 257 40258 R Raw gas sensor data ADC data 0...

Страница 6: ...sure to high levels of ethyl acetate results first in irritation of the eyes nose and throat followed by headache nausea vomiting sleepiness and unconsciousness Exposure limits NIOSH TWA 8 hours 1400...

Страница 7: ...Semi Volatile Organic Compound SVOC cannot be measured NA Not available Occasionally you will be measuring a mixture of VOCs If the total concentration is within the linear range of your PID then it i...

Страница 8: ...29 Benzyl acetate C9H10O2 140 11 4 9 0 6 30 Benzyl alcohol C7H8O 100 51 6 8 26 1 6 125 31 Camphor C10H16O 76 22 2 8 76 0 4 32 Carbon dioxide CO2 124 38 9 13 77 NR 0 0 33 Carbon disulfide CS2 75 15 0 1...

Страница 9: ...hyl benzoate C9H10O2 93 89 0 8 9 0 9 68 Ethyl butyrate C6H12O2 105 54 4 9 9 1 4 5 100 69 Ethylamine C2H7N 75 04 7 8 86 1 5 100 70 Ethylbenzene C8H10 100 41 4 8 76 0 56 3 50 71 Ethylcyclohexane C8H16 1...

Страница 10: ...C8H18O 26952 21 6 9 8 1 7 9 170 143 Isopentane C5H12 78 78 4 10 32 4 30 600 144 Isopentanol C5H12O 137 32 6 9 86 2 145 Isopentene C5H10 563 46 2 9 12 0 8 146 Isoprene 2 methyl 1 3 butadiene C5H8 78 79...

Страница 11: ...alic acid C2H2O4 144 62 7 11 2 NR 182 Oxalyl bromide C2Br2O2 15219 34 8 10 49 5 183 Oxydiethanol 2 2 diethylene glycol C4H10O3 111 46 6 10 3 2 20 400 184 Oxygen O2 7782 44 7 12 07 NR 185 Ozone O3 1002...

Страница 12: ...8 9 6 0 7 221 Rose oxide cis Tetrahydro 4 methyl 2 2 methylpropenyl 2H pyran C10H18O 16409 43 1 9 0 8 222 Sec amyl acetate C7H14O2 626 38 0 9 9 5 223 Sevoflurane 1 1 1 3 3 3 hexafluoro 2 fluoromethox...

Страница 13: ...window that presents itself as a blue hue To check for contamination hold the lamp in front of a light source and look across the window surface To avoid contaminating the sensor and affecting accurac...