background image

E2648-PID User Manual
Annex 4 Calibration and maintenance

Do not perform any maintenance operation with the power on.

Clean the device with soft damp cloth. Do not use any abrasive cleaning agents. Do
not immerse the device into water or any cleaning media.

NOTE

  Cleaning  instruments  and  spare  parts  are  not  included  to the  delivery  set,

please contact your Seller for more information.

How often the PID needs maintenance

The  frequency  of  calibration  and  maintenance  depends  on  operating  conditions
(VOC concentration, humidity, dust and oil in the air etc).
In  clean  indoor  air  with  low  VOC  concentration  a  monthly  or  less  frequent
calibration is recommended. When higher VOC concentrations are measured under
harsh conditions, perform calibration and maintenance more frequently.
The PID needs maintenance:

If the baseline is climbing after you zero the PID – replace electrode stack
If the PID becomes sensitive to humidity – replace electrode stack
If  the  PID  cell  shows  signs  of  contamination  after  the  lamp  window  has  been
cleaned – replace electrode stack
If the PID cell is known to have been subjected to severe contamination– replace
electrode stack
If the baseline is unstable or shifts when you move the PID – replace electrode
stack
If  sensitivity  has  dropped  too  much  (note  the  change  required  when  checking
calibration) – clean the lamp
if  the  signal  to  noise  ratio  at  low  VOC  concentrations  becomes  inadequate  –
replace the lamp

Cleaning  of  the  PID  lamp  is  recommended  as  the  first  action  when  a  PID  needs
maintenance.  It  is  recommended  to  recalibrate  the  device  every  time  after  stack
replacement or lamp cleaning.

Removing the electrode stack and lamp

Always  use  the  Electrode  Stack  Removal  Tool  to  remove  the  electrode
stack; any other tools may damage your PID and invalidate your warranty.

To remove the electrode stack, proceed as follows:
1. Gently remove the sensor from equipment.
2. Place the PID, pellet side down, onto a clean surface.
3. Locate electrode stack removal tool into the two slots on the sides of the PID and
squeeze together until electrode stack and lamp are released.
4. Carefully lift the PID body away from the pellet and lamp.
5. Occasionally the lamp may be temporarily lodged in the cell and will need to be
freed carefully with tweezers.
6. Occasionally  the  small  spring  behind  the  lamp  will  come  out  when  the  lamp  is
removed from the sensor. Simply replace it into the sensor house.

Cleaning the PID Lamp

Always use recommended cleaning kit to clean the lamp.

Inspection of the lamp may reveal a layer of contamination on the detection window
that  presents  itself  as  a  'blue  hue.'  To  check  for  contamination,  hold  the  lamp  in
front of a light source and look across the window surface To avoid contaminating
the sensor and affecting accuracy, do not touch the lamp window with bare fingers.
You may touch the body of the lamp with clean fingers.

PID lamp cleaning kit

The  vial  of  cleaning  compound  contains  alumina  (CAS  Number  1344-28-1)  as  a
very  fine  powder.  Please  contact  your  Seller  for  full  material  safety  data  sheet
MSDS.
Key safety issues are identified below:
Hazard identification: May cause irritation of respiratory tract and eyes
Storage: Keep container closed to prevent water adsorption and contamination
Handling:
• Do not breathe in the powder. Avoid contact with skin, eyes and clothing
• Wear suitable protective clothing
•  Follow  industrial  hygiene  practices:  Wash  face  and  hands  thoroughly  with  soap
and water after use and before eating, drinking, smoking or applying cosmetics
• The powder carries a TVL(TWA) limit of 10 mg/m

3

PID lamp cleaning kit

1. Open the container of alumina polishing compound.
2. With a clean cotton bud, collect a small amount of the powder.
3.  Use  this  cotton  bud  to  polish  the  PID  lamp  window.  Use  a  circular  action,
applying  light  pressure  to  clean  the  lamp  window.  Do  not  touch  the  lamp  window
with fingers.
4.  Continue  polishing  until  an  audible  “squeaking”  is  made  by  the  cotton  bud
moving over the window surface. (usually within 15 seconds)
5. Remove the residual powder from the lamp window with a clean cotton bud. Care
must be taken not to touch the tips of cotton buds that are to be used to clean the
lamps as this may contaminate them with finger oil.
6. Ensure the lamp is completely dry and remove any visible signs of contamination
before refitting.

Re-fitting the PID electrode stack and lamp

WARNING

: Never refit a damaged lamp.

1. Place the lamp inside the O-ring seal in the pellet as illustrated below.
Twisting  the  lamp  slightly  during  insertion  will  help  to  ensure  the  lamp  window  is
snug  against  the  electrode  stack’s  front  electrode.  The  lamp  should  be  freely
supported by the O-ring.
2.  Continuing  to  hold  the  electrode  stack between  forefinger  and  thumb,  carefully
insert the lamp into recess in the sensor ensuring that the lamp remains in position.
Press the  electrode stack firmly, to ensure that the  electrode stack wing  clips are
engaged, and the top faces of the electrode stack and sensor house are flush.
3. Refit the sensor into the sensing equipment.
4. Recalibrate the gas detector.

Discarding the PID electrode stack

The electrode stack does not have any toxic components, however, if it has been
contaminated by toxic materials, show due care when disposing.

Feel free to contact us if you have questions.

Electrode stack

Electrode stack removing tool

Содержание PluraSens E2648-PID

Страница 1: ...erial and workmanship for a period of one year from the date of original sale During this warranty period Manufacturer will at its option either repair or replace product that proves to be defective T...

Страница 2: ...and or digital interface terminals to the necessary devices according to the relevant connection diagram The screwless quick connect spring terminals on the E2648 series devices are suitable for a wi...

Страница 3: ...may be addressed either from 00001 or 40001 base Addr Reg MHR RW Description Supported values dec Default 0x0001 1 40002 R Hardware version 0x0002 2 40003 R Software version 0x0003 3 40004 R Product s...

Страница 4: ...in MHR 40205 2 0x00CB 203 40204 RW Forced value for analog output 1 0 1000 0 0 100 0 of output scale 0 0x00CC 204 40205 RW Forced value for analog output 2 0 1000 0 0 100 0 of output scale 0 0x00D3 2...

Страница 5: ...0 1 LED is on off in normal condition bit 11 0 1 LED and buzzer syncing off on with RE1 bit 12 0 1 LED and buzzer syncing off on with RE2 user defined 0x0101 257 40258 R Raw gas sensor data ADC data 0...

Страница 6: ...sure to high levels of ethyl acetate results first in irritation of the eyes nose and throat followed by headache nausea vomiting sleepiness and unconsciousness Exposure limits NIOSH TWA 8 hours 1400...

Страница 7: ...Semi Volatile Organic Compound SVOC cannot be measured NA Not available Occasionally you will be measuring a mixture of VOCs If the total concentration is within the linear range of your PID then it i...

Страница 8: ...29 Benzyl acetate C9H10O2 140 11 4 9 0 6 30 Benzyl alcohol C7H8O 100 51 6 8 26 1 6 125 31 Camphor C10H16O 76 22 2 8 76 0 4 32 Carbon dioxide CO2 124 38 9 13 77 NR 0 0 33 Carbon disulfide CS2 75 15 0 1...

Страница 9: ...hyl benzoate C9H10O2 93 89 0 8 9 0 9 68 Ethyl butyrate C6H12O2 105 54 4 9 9 1 4 5 100 69 Ethylamine C2H7N 75 04 7 8 86 1 5 100 70 Ethylbenzene C8H10 100 41 4 8 76 0 56 3 50 71 Ethylcyclohexane C8H16 1...

Страница 10: ...C8H18O 26952 21 6 9 8 1 7 9 170 143 Isopentane C5H12 78 78 4 10 32 4 30 600 144 Isopentanol C5H12O 137 32 6 9 86 2 145 Isopentene C5H10 563 46 2 9 12 0 8 146 Isoprene 2 methyl 1 3 butadiene C5H8 78 79...

Страница 11: ...alic acid C2H2O4 144 62 7 11 2 NR 182 Oxalyl bromide C2Br2O2 15219 34 8 10 49 5 183 Oxydiethanol 2 2 diethylene glycol C4H10O3 111 46 6 10 3 2 20 400 184 Oxygen O2 7782 44 7 12 07 NR 185 Ozone O3 1002...

Страница 12: ...8 9 6 0 7 221 Rose oxide cis Tetrahydro 4 methyl 2 2 methylpropenyl 2H pyran C10H18O 16409 43 1 9 0 8 222 Sec amyl acetate C7H14O2 626 38 0 9 9 5 223 Sevoflurane 1 1 1 3 3 3 hexafluoro 2 fluoromethox...

Страница 13: ...window that presents itself as a blue hue To check for contamination hold the lamp in front of a light source and look across the window surface To avoid contaminating the sensor and affecting accurac...

Отзывы: