Page 35
HA249928
Issue 15 Jly 06
100 MM CHART RECORDER USER GUIDE
150mV
1 Volt
10 Volts
-30 mV
-0.2 Volt
-2 Volts
5.5µV
37µV
370µV
0.084% input + 0.053% range
0.084% input + 0.037% range
0.275% input + 0.040% range
80ppm of input per deg C
80ppm of input per deg C
272ppm of input per deg C
Low
Range
High
Range
Resolution
Maximum error
(Instrument at 20˚C)
Worst case temperature
performance
B
C
D
E
G2
J
K
L
N
R
S
T
U
Ni/NiMo
Platinel
0 to + 1820
0 to + 2300
0 to + 2495
- 270 to + 1000
0 to + 2315
- 210 to + 1200
- 270 to + 1372
- 200 to + 900
- 270 to + 1300
- 50 to + 1768
- 50 to + 1768
- 270 to + 400
- 200 to + 600
0 to + 1406
0 to + 1370
IEC 584.1
Hoskins
Hoskins
IEC 584.1
Hoskins
IEC 584.1
IEC 584.1
DIN43700:1985
(To IPTS68)
IEC 584.1
IEC 584.1
IEC 584.1
IEC 584.1
DIN 43710:1985
Ipsen
Engelhard
T/C
Type
Overall range
(˚C)
Standard
0 to 400˚C: 1.7˚C
400 to 1820˚C: 0.03˚C
0.12˚C
0.08˚C
0.03˚C
0.07˚C
0.02˚C
0.04˚C
0.20˚C
0.04˚C
0.04˚C
0.04˚C
0.02˚C
0.08˚C
0.14˚C
0.02˚C
Max linearisation error
6.2 TECHNICAL SPECIFICATION (Input board)
General
Termination
Edge connector / terminal block
Maximum number of inputs
Continuous trace recorder = 4. Multipoint recorder = 6.
Input ranges
- 30 to + 150 mV; - 0.2 to + 1 Volt; - 2 to + 10 Volts;
Input types
dc Volts, dc millivolts, dc milliamps (with shunt), thermocouple, 2 / 3-wire resistance temperature detector (RTD)
(Channel 1 can be RTD or Ohm only if no other channel is thermocouple - see section 3.2.9 Note 1)
)
Input type mix
User configurable
Noise rejection (48 to 62 Hz)
Common mode: >140dB (channel - channel and channel - to - ground). Series mode: >60dB.
Maximum common mode voltage
250 Volts continuous
Maximum series mode voltage
180 mV at lowest range; 12 Volts peak at highest range.
Isolation (dc to 65 Hz; BS EN61010)
Installation category II; Pollution degree 2 (See section 5 above for definitions)
300V RMS or dc Channel to channel (double insulation), channel to common electronics (double insulation) and
channel to ground (basic insulation)
Dielectric strength (BS EN61010)
Channel to ground =1350 Vac; Channel to channel = 2300 Vac (both 1 minute type tests)
Insulation resistance
>10M
Ω
at 500 V dc
Input impedance
150 mV and 1 V ranges: >10 M
Ω
; 10 V range: 68.8 k
Ω
Over voltage protection
50 Volts peak (150V with attenuator)
Open circuit detection
±
57 nA max.
Recognition time
Continuous trace = 250 msec; Multipoint = 500 msec
Minimum break resistance
10 M
Ω
DC Input ranges
Shunt/attenuator
Externally mounted resistor modules
Additional error due to shunt
0.1% of input
Additional error due to attenuator
0.2% of input
Performance
Thermocouple data
Temperature scale
ITS 90
Linearisation accuracy
0.05% of user entered span
Bias current
0.05 nA
Cold junction types
Off, internal, external
CJ error
1˚C max; instrument at 25 ˚C
CJ rejection ratio
50:1 minimum
Upscale / downscale drive
High, low or none
Types and ranges
Содержание 4102C
Страница 1: ...4102C M 100mm chart recorders User Guide EUROT HERM ...
Страница 4: ......
Страница 44: ...HA249928 Issue 15 Jly 06 Page 40 100 MM RECORDER USER GUIDE This page is deliberately left blank ...
Страница 48: ......
Страница 49: ......
Страница 50: ......
Страница 51: ......
Страница 52: ......