36 | EUROMAG
Error
code
J
H
Q
F
1
2
3
K
Message
on window M08
Hardware Error
Poor Sig. Detected
Frequency Output Over
System RAM Error
Date Time Error
CPU or IRQ Error
ROM Parity Error
Adjusting Gain
Empty pipe
Causes
Hardware problem
1)Poor signal detected
2)Transducers installed improperly
3)Too much fouling (corrosion,
deposition, etc.)
4)The pipe liner is too thick.
5)Problem with transducer cables
The actual frequency for the
Frequency Output is out of the
range specified by the user
1) Temporary problems with RAM,
RTC
2) Permanent problems with
hardware
Instrument is in the progress of
adjusting the gain for the signal,
and the number indicates the
progressive steps
No liquid inside the pipe
Incorrect setup in M29
Solutions
Contact the manufacturer
1)Adjust measuring location
2)Polish the pipe surface and clean the
spot
3)Make sure the couplant is enough
4)Check the transducer cables
Check the values entered in window
M66, M67, M68 and M69, and use a
larger value in M69
1) Turn on the power again
2) contact the manufacturer
No need for action
Relocate the meter to where the pipe is
full of liquid
Enter 0 in M29
It comes from the previous page
EUROSONIC 2000 HH
Содержание EUROSONIC 2000 HH
Страница 2: ...EUROSONIC 2000 HH...
Страница 6: ...6 EUROMAG EUROSONIC 2000 HH...
Страница 12: ...12 EUROMAG EUROSONIC 2000 HH...
Страница 28: ...28 EUROMAG EUROSONIC 2000 HH...
Страница 34: ...34 EUROMAG EUROSONIC 2000 HH...
Страница 38: ...38 EUROMAG EUROSONIC 2000 HH...
Страница 44: ...44 EUROMAG EUROSONIC 2000 HH...
Страница 77: ...EUROMAG 77...
Страница 78: ...78 EUROMAG EUROSONIC 2000 HH...
Страница 79: ...EUROMAG 79...