EUROMAG | 33
MENU WINDOWS DETAILS
Menu window No.
M85
M86
M87
M88
M89
M90
M91
M92
M93
M94
M95
M96
M97
M98
M99
M+0
M+1
M+2
M+3
M+4
M+5
M+6
M+7
M+8
M+9
M-0
Function
Not used
Not used
Select transducer power between 1-10 (default 10)
Set receive window start
Set receive window end
Display signal strength, signal quality and transit time ratio (upper right corner).
Display the transit time ratio. The ratio value should be in the range of 100±3% if the
entered pipe parameters are correct and the transducers are properly installed. Otherwise,
the pipe parameters and the transducer installation should be checked.
Display the estimated sound speed of the fluid in the pipe. If this value has an obvious
difference with the actual fluid sound speed, the user is recommended to check if the pipe
parameters are correct and if the transducer installation is good.
Display the total transit time and delta time (transit time difference between upstream and
downstream travelling)
Display the Reynolds number and the pipe factor used by the flow rate measurement
program. Note, the pipe factor is rarely used.
Not used
Not used
Command to store the pipe parameters either in the built-in data logger or to the RS-232C
serial interface
Command to store the diagnostic information either in the built-in data logger or to the
RS-232C serial interface
Command to copy the current display either to the built-in data logger or to the RS-232C
serial interface
View the last 64 records of power on and off events. The recorded information include the
date and time as well as the corresponding flow rate when the power on or off occurs
Display the total working time of the instrument
Display the last power-off date and time
Display the last power-off flow rate
Display the total number of times the flow meter has been powered on and off
A scientific calculator for the convenience of field applications.
All the values are in single accuracy.
All the mathematic operators are selected from a list.
Velocity change setup
Protocol selection
Not used
Not used
Entry to hardware adjusting windows. Valid for the manufacturer only.
Содержание EUROSONIC 2000 HH
Страница 2: ...EUROSONIC 2000 HH...
Страница 6: ...6 EUROMAG EUROSONIC 2000 HH...
Страница 12: ...12 EUROMAG EUROSONIC 2000 HH...
Страница 28: ...28 EUROMAG EUROSONIC 2000 HH...
Страница 34: ...34 EUROMAG EUROSONIC 2000 HH...
Страница 38: ...38 EUROMAG EUROSONIC 2000 HH...
Страница 44: ...44 EUROMAG EUROSONIC 2000 HH...
Страница 77: ...EUROMAG 77...
Страница 78: ...78 EUROMAG EUROSONIC 2000 HH...
Страница 79: ...EUROMAG 79...