INSTALLATION
Install
the instrument away from heat sources in a dry place
at a maximum room temperature of 40°C.
Comply with standards in force in the different countries
regarding electrical installation (Fig. 2).
If the power cord does
not have a plug, the device must be connected to the mains
by means of an omnipolar disconnecting switch with at least
3 mm between the contacts. All the power supply circuits
must be interrupted before accessing the connection
devices
.
Fig. 1
–
Electrical connection
W al l i nstal l ati on
The wall-mounting plugs are supplied with the device.
Always use a plug suitable to the available support.
The layout of the
holes to be drilled on the support is displayed in Figure 2.
Fig. 2
-
Measurements for wall fixing 226l x 146h
To access the 4 installation holes, remove the covers on the installation points
(A)
found on each corner of the
instrument, use a Phillips screwdriver to loosen the four screws underneath the covers, then open the front panel (see Fig.2).
The casing has 4 captive screws to quickly open/close the cover, thereby allowing for easy access for commissioning
and servicing, as well as assuring excellent seal for long-lasting operation
Install the unit in a dry place away from heat sources. Max room temperature 40°C.
Strictly comply with the regulations in force in the various countries regarding electrical systems.
Fit the instrument on the wall using the screws supplied
D i ag ram of el ectri cal connecti ons
To connect the accessories and peripheral devices to the instrument, remove the front cover screws, using a Phillips screwdriver
to reach the connecting terminal boards.
The terminal boards consist of spring terminals for qui
ck coupling of the wires. Press the square “slotted” pin with a small flat
headed screwdriver and insert the stripped wire in the corresponding terminal.
ATTENTION: exert slight pressure on the spring
pin to avoid irreparably damaging the terminal board.
Do not connect more than one device to each pin
Run the wires to be connected through the cable glands on the case wall.
Covers
(A)
x 4