Modular pulses, also known as PWM “pulse width modulation”, support a proportional mode on each ON
-OFF
setpoint, activating the corresponding pulses on the output relays based on the measured value and always
relating to the previous selected settings.
The default activation point is
0.050 ppm
(2 Cl ppm);
0.50 ppm
(range 20 Cl ppm, 0-10ppm CLC Cell).
For more accurate settings of the
“Modular pulses”
PWM, select the
“Expert”
menu in the initial settings.
*
GOING BACK TO THE SETPOINT PROGRAMMING STEPS FROM
“ON
-
OFF”
MODE OR FROM
“PROPORTIONAL”
MODE
FUNCTION ONLY AVAILABLE WITH SETPOINT 1
The MIN and MAX Alarm function selects the alarm levels beyond which the alarm relay is triggered on.
ONLY WHEN USING METERING PUMPS
The priming function blocks the setpoint value to allow the metering pump to
remain primed, the outputs are not active.
Press
E S C
to go back to
M E N U S E L E C T I O N
or press
E S C
E S C
to go back to the
M E A S U R E M E N T
D I S P L A Y
A
CALIBRATION > BASIC MENU
◄
►
Press
O
O K / M E N U
U
to confirm the selection and go to the next sub-menu.
The chlorine parameter has 3 ranges suitable for
ion-selective membrane
sensors
(one of which for the Total chlorine) and a range
for open type
amperometric cells
.
NOTE
During calibration, the
-
(minus) sign might be displayed, which disappears or is removed by correct settings! The value in the lower part of
the display shows the sensor signal in mV thus indicating the sensor’s efficiency. It is advisable to perform constant checks
and periodic
recalibration of the sensor, via DPD1.
For the calibration
of
ion-selective membrane
sensors always refer to the probe’s manual.
In BASIC MENU mode calibration is performed with one point only, the DPD one.
Select the value of the required sample solution and check with the
DPD1
test kit for free chlorine, or with portable instrument, measure the
chlorine concentration in ppm (residual or total) in water, wait for the measurement to stabilise, p
ress and hold ▲▼ until showing the value
obtained with the DPD method then confirm with
O
O K / M E N U
U
MENU SETPOINT 1 -2
Mode
Prop ortion al
MENU BASE
MENU SETPOINT 1
Alarm MIN
0.000 Cl ppm
MENU BASE
MENU SETPOINT 1
Alarm MIN
0.00 Cl ppm
MENU BASE
MENU SETPOINT 1
Alarm MIN
0.00 Cl ppm
* (Tot )
MENU BASE
MENU SETPOINT 1
Alarm MIN
0.00 Cl p p m
MENU BASE
MENU SETPOINT 1
Alarm MAX
2,000 Cl p p m
MENU BASE
MENU SETPOINT 1
Alarm MAX
20.00 Cl p p m
MENU BASE
MENU SETPOINT 1
Alarm MAX
20.00
Cl
p p m
* ( T ot )
MENU BASE
MENU SETPOINT 1
Alarm MAX
10.00 Cl p p m
MENU BASE
MENU SETPOINT 1-2
Priming
OF F
MENU BASE
MENU SETPOINT 1-2
Priming
ON
MENU BASE
Menu Selection
Calib r ati on
MEASUREMENT
Calibrat.
0.000 Cl p p m
42.1 mV
MEASUREMENT
Calibrat.
0.00 Cl p p m
42.1 mV
MEASUREMENT
Calibrat.
0.00
Cl
p p m * ( T ot )
420.1 mV
MEASUREMENT
Calibrat.
CLC 0.00 C l
p p m
420.1 mV