Maintenance
Cleaning and disinfection
Clean on a regular basis. Clean surfaces with
a damp cloth using an appropriate pH-neutral
detergent. Do not use solvents or strong liquids,
this may damage surfaces on the lifter. For
disinfection when needed; use isopropyl alcohol.
Avoid abrasive cleaning products. Remove hair
and pile from the castors and verify that the
castors rotates freely. Clean contact and hand
control with isopropyl alcohol to remove grease
and dirt. Check emergency stop and emergency
lowering after cleaning. The lift should not be
exposed to running water.
Make sure not to damage or
remove labels when cleaning.
Reconditioning
Follow cleaning and assembly procedure, complete
periodic inspection and use checklist after assem-
bly to recondition the lifter.
Recycling
Refer to “Recycling instructions” for how to
properly dispose of product. This can be found on
www.etac.com.
Spare Parts
A list of spare parts is available on request.
Periodic inspection
Periodic inspection scope
Periodic inspection is a visual examination (particu-
larly of the lifter’s load bearing structure and lifting
mechanism with attachments, brakes, controls,
safety devices and person-support devices) accord-
ing to Periodic Inspection Report for Molift Mover
300. This can be found on www.etac.com.
If there is any problems with the
lifter that could jeopardize some-
ones safety, the lifter shall imme-
diately be taken out of service and
marked “out of order”. Do not use
the lifter untill it is repaired.
Periodic Inspection shall be performed annually or
more frequently if required by local requirements.
The inspection must be performed by service
personnel authorized by Etac.
Contact Etac at [email protected] for training and
authorization or recommendation of an approved
service partner.
When performing a periodic inspection, the
inspector shall fill out the inspection report for
Molift Mover 300. The reports should be retained
by the person(s) responsible for servicing the lifter.
If the inspection reveals defects and damages, the
owner shall be notified.
When periodic inspection is completed the
inspector shall mark the lifter with a sticker on
the control label showing the date when periodic
inspection is performed and a label (with 6-digit
ID from Etac or distributor) showing who has
performed the periodic control. The control label
can be found on the right side of the battery
holder, and this will then indicate the date when
next Inspection is due.
Mark label
with month
and year of
inspection
16
Molift Smart 150 / www.etac.com
Содержание Molift Mover 300
Страница 2: ......
Страница 3: ...Molift Mover 300 EN User manual BM06101 Rev G 2021 02 16...
Страница 18: ......
Страница 19: ...Molift Mover 300 DE Bedienungsanleitung BM06102 Rev E 2021 02 16...
Страница 34: ......
Страница 35: ...Molift Mover 300 NO Brukermanual BM06103 Rev E 2021 02 16...
Страница 50: ......
Страница 51: ...Molift Mover 300 SV Bruksanvisning BM06104 Rev E 2021 02 16...
Страница 66: ......
Страница 67: ...Molift Mover 300 FI K ytt ohje BM06105 Rev E 2021 02 16...
Страница 82: ......
Страница 83: ...Molift Mover 300 IT Manuale dell utente BM06106 Rev B 2021 02 16...
Страница 98: ......
Страница 99: ...Molift Mover 300 FR Manuel d utilisation BM06107 Rev E 2021 02 16...
Страница 114: ......
Страница 115: ...Molift Mover 300 ES Manual de usuario BM06108 Rev L 2021 02 16...
Страница 130: ......
Страница 131: ...Molift Mover 300 NL Gebruikershandleiding BM06109 Rev E 2021 02 16...
Страница 146: ......
Страница 147: ...Molift Mover 300 DA Brugervejledning BM06110 Rev E 2021 02 16...
Страница 162: ......
Страница 163: ......
Страница 164: ...Etac A S Parallelvej 3 DK 8751 Gedved www etac com...