20
Red level led blinks in different ways described in the table
Tab. 4. Led LEVEL
LED
STATE
SOLUTION
Permanent red
Product end (if present a level
probe) / tank empty
Fill the tank
3 blinks per
second
Power supply is over the
range (refer to pump label)
Check power supply correspond to pump
label. Shutdown and restart.
2 blinks per
second
Power supply is under the
range (refer to pump label)
Check power supply correspond to pump
label. Shutdown and restart.
1 blink per
second
Pump is waiting program
setting
Press scroll key and choose a program.
Confirm with P key
LEVEL led
PROGRAMS led shows the current working program.
Press repeatedly SCROLL to select hte working program
Tab. 5. Led PROGRAMS
LED
STATE
Permanent on
Pump ON. Current pump working mode.
1 blink every 2
seconds on last
working program.
Pump OFF..
All leds blinking
together
Pump is waiting for programming. Press P and SCROLL to select the
program or wait 30 seconds to exit without changing.
PROGRAMS led