Operation Summary
Care@Home
™
PERS Control Panel User Guide
15
Figure 8 – Securing the Power Adapter Cube
NOTE:
The Bracket must be installed to maintain the UL Home Healthcare (UL 1637)
Listing. The Bracket may not be used for the ULC Canadian Residential Fire Alarm (UL
S545) Listing.
1.
Turn the Power Switch ON
2.
Return the bottom cover back to its place.
There is an option to secure it with a screw.
3.
Install the Control Panel (CP) in its final location.
Once the power is switched ON, the CP goes into standby mode, as indicated by the Power Indicator turning
green and the Communication Indicator turning red.
If the CP was delivered pre-configured, the ring around the
EMERGENCY
button turns blue as well.
NOTE:
When testing the system, test with the AC connected and then with the AC
disconnected. If there are restraints, remove the restraints during testing. Reconnect
the restraints when testing is completed.
Secure the bracket with a
screw for standard style
outlets
Power Adapter
Secure the bracket with a screw
for decorative style outlets