PLATINOUS SERIES CONTROLLER MANUAL
43
Absolute high/low limits
:
These settings are designed to protect the
chamber against thermal damage. Set them at
least 10°C higher/lower than the target
temperature. When tripped, an alarm is
generated and the chamber stops running.
Upper deviation limit
:
This setting is designed to protect specimens
against heat damage. Basically, it determines the
temperature at which the chamber heater is shut
OFF. Set it according to the type of specimen
tested. When tripped, an alarm is generated and
the chamber stops running. However, when
temperature returns within range, normal control
is restored automatically.
5.4. Sampling Setup and ON/OFF
1.
Get the chamber setup mode.
Press the Chamber Setup box on the main menu.
TEMP.
2.
Press the Sampling Setup box on the Chamber Setup screen.
Содержание SCP-220
Страница 1: ...SCP 220 CONTROLLER INSTRUCTION MANUAL...
Страница 2: ......
Страница 38: ...PLATINOUS SERIES CONTROLLER MANUAL 34...
Страница 69: ...PLATINOUS SERIES CONTROLLER MANUAL 65...
Страница 79: ...PLATINOUS SERIES CONTROLLER MANUAL 75...