100
A-3 General Specifications
General Items
Specifications
Warm up time
At least 30 minutes
Temperature Coefficient
Add 0.15 x (the applicable accuracy)/
°
C at 0
°
C to 18
°
C and
28
°
C to 50
°
C
Operating Temperature
0
°
C to 50
°
C (32
°
F to 122
°
F)
Storage Temperature
-20
°
C to 60
°
C
Altitude
Up to 2000 M
Pollution Degree
II
Over-voltage Category
CAT III-600V and CATII-1000V
Relative Humidity
Up to 80% (60% for 50M
Ω
range of resistance measurement)
Common Mode Voltage
1000V dc or peak ac rms maximum between any input and
earth ground
Dimension
Approx. 255(w) x 105(h) x 305(d) mm (with holsters)
Weight <3.0kgs
Line Voltage
100V / 120V / 220V / 240V ac
±
10%, 50/60Hz, 16VA maximum
Interface
RS-232 (DB-9, male connector)
Baud rates: 9600, 4800, 2400, 1200, 600, 300
Data length: 7 or 8 bits
Parity: even / odd / none
Stop bit: 1 or 2 bits
Echo: on / off
Print mode: on / off
Safety Requirement
Designed in compliance with EN61010-1 (IEC1010-1)
Installation Category
CAT-II 750VAC/1000VDC or CAT-III 600V, Pollution Degree
II Environment
EMC Requirement
Designed in compliance with EN61326-1.