52
Query Commands
The purpose of
Query
commands is used for requesting the meter to respond
its current status. An example of a query command <
R1
> is used for
requesting the meter to respond its primary display characters.
Please refer to Section 6-5-3 for detail information about Query Commands.
6-4-2 Command Syntax
Echo
With echo ON, the meter echoes (returns) all the characters whatever it
receives.
Terminator
A terminator is a character sent by a host, which identifies the end of a
command string. A valid terminator consists of two-byte data:
<CR> (Carriage Return) and <LF> (Line Feed)
Prompts
When a host sends a command string to the meter through RS-232 interface,
the meter executes the command and returns one of the prompts as shown
on Table 6-2.
Table 6-2. RS-232 Return Prompts
Prompts
Description
﹡
>
The meter is reset to power-up initialisation status.
= >
A command is executed and no errors are detected.
! >
A command error is detected.
? >
A parameter error is detected.
# >
The local key is pressed.
S >
The set up function is under executing.
@ >
No numeric reading is available.
W>
When the setting value HI<LO or LO>HI.
E>
Execution error, or disallow to execute. For example, the
K13 will not be allowed after LLO.
Return result
After the meter executes a query command the return of the result will be in the
following format:
<RESULT> + <CR> <LF> + <PROMPT> + <CR><LF>