Problem 8
—
Lights always OFF
PROBLEM
POSSIBLE CAUSE
CORRECTIVE ACTION
Lights always OFF
Faulty fluorescent tube
•
Replace the faulty fluorescent tube.
•
Fluorescent tube is located inside the blue panel (Refer to
Fig. 7.3 for lamp location and replacement).
Faulty Fluorescent ballast
For PCR:
•
See Layout A at the end of this section to locate the
fluorescentballast, connector B, and relay board inside
electricalpanel.
•
Turn on the cabinet by connecting to the main supply, then
turn on the light by pressing LIGHT button on
membrane/keypad.
•
Check AC voltage at ballast input (between pin NO on J13
terminal on relay board to neutral), see Fig. 8.1below.
•
It should be 230 VAC ± 10% for 230 V cabinets or 115 VAC
± 10% for 115 V cabinets.
•
If not, check the LS7 relay and F7 fuse (refer to possible
cause
Faulty relay or fuse
below).
For SCR:
•
See Layout B at the end of this section to locate the
fluorescent ballast and connector B inside electrical panel.
•
Turn on the cabinet by connecting to the main supply, then
switch on the light with the LIGHT rocker switch on control
panel.
•
Check AC voltage at ballast input (between pin connector
A-5 to neutral). It should be 230 VAC ± 10% for 230 V
cabinets or 115 VAC ± 10% for 115 V cabinets.
Fig 8.1
DOMINIQUE
DUTSCHER
SAS
Содержание Airstream PCR-3A
Страница 6: ...D O M I N I Q U E D U T S C H E R S A S...
Страница 13: ...D O M I N I Q U E D U T S C H E R S A S...
Страница 14: ...Polymerase Chain Reaction Cabinets 2 1 1 Certification Flowchart D O M I N I Q U E D U T S C H E R S A S...
Страница 34: ...D O M I N I Q U E D U T S C H E R S A S...
Страница 57: ...44 Polymerase Chain Reaction Cabinets LAYOUT A D O M I N I Q U E D U T S C H E R S A S...
Страница 58: ...45 LAYOUT B D O M I N I Q U E D U T S C H E R S A S...
Страница 60: ...47 D O M I N I Q U E D U T S C H E R S A S...
Страница 61: ...48 APPENDIX D O M I N I Q U E D U T S C H E R S A S...