![Esco Airstream PCR-3A Скачать руководство пользователя страница 19](http://html1.mh-extra.com/html/esco/airstream-pcr-3a/airstream-pcr-3a_service-manual_2426103019.webp)
11
1
5
Cabinet Model
3 feet
4 feet
Width =
w
(mm) (in)
930 (36.6)
1235 (48.6)
Depth =
h
(mm) (in)
530 (20.9)
530 (20.9)
Left to
Right
Distance from walls =
a
(mm) (in)
300 (11.8)
300 (11.8)
Distance apart =
x
(mm) (in)
330 (12.9)
635 (25)
Front to
Back
Distance from walls =
b
(mm) (in)
265 (10.4)
265 (10.4)
Average light reading:
lux (ft.-candles)
Acceptance:
Average light reading should be at least 800 lux (75 ft-candles)
Results:
Pass / Fail
Acceptance criteria / testing procedures have been adopted in accordance with the requirements of the: IEST-
RP-CC002.3 Section 6.10 and AS 1807.15 “Illuminance (Lightning Intensity)”