C U S T O M I Z I N G M Y L A B
MyLab
- G E T T I N G S T A R T E D
6 - 9
GETTI
NG STARTED
Corrupted System Profile
Whenever a profile is corrupted, a red exclamation mark is displayed on the Profile
Manager option. In this case, enter the profile manager menu and, for each system
profile, check that every component has a specific setting (no component has to
be without any setting).
Should this not solve the problem, contact Esaote personnel.
Center ID
Center ID allows to set the center name displayed in the Heading Area of the
screen and the center information shown in the report.
To configure the Center ID press
M EN U
, then select
CE NT ER ID
and then
follow the generic configuration procedure.
During editing you can configure many fields described below.
Center ID Field
The name entered in this field will be displayed in the heading area of the screen.
Report Information Field
This option allows to add to the report header the following information:
the hospital name;
the department name;
the contact details;
two fields for other additional information;
the hospital logo.
DICOM Field
This option allows to enter the station name used in DICOM.
General Settings
AO
This chapter explains how to set many
MyLab
options. For configurations not
described here refer to relevant chapters in the “Advanced Operations” manual.
NOTE
Some settings described below may be unavailable on your product’s
configuration.
Содержание MyLabX75VET
Страница 1: ...Rev 03 June 2021 MyLabX75VET GETTING STARTED 350080800 ...
Страница 2: ......
Страница 5: ...MyLab G E T T I N G S T A R T E D v GETTING STARTED RED Declaration of Conformity ...
Страница 6: ...MyLab G E T T I N G S T A R T E D vi ...
Страница 61: ...S Y S T E M O V E R V I E W MyLab G E T T I N G S T A R T E D 3 17 GETTING STARTED ...