![Ergotron StyleView SV32 Скачать руководство пользователя страница 19](http://html1.mh-extra.com/html/ergotron/styleview-sv32/styleview-sv32_user-manual_2422613019.webp)
19/21
EN
888-24-092-G-02 rev. H • 05/11
Maintenance & Safety
Cleaning and Maintenance
The following procedures are not guaranteed to control infection. The hospital infection control administrator or epidemiologist should be consulted regarding cleaning procedures
and processes.
To avoid risk of electric shock, do not expose electrical components to water, cleaning solutions or other potentially corrosive liquids or substances.
Do not immerse Cart or Cart components in liquid or allow liquids to fl ow into the Cart. Wipe all cleaners off surface immediately using a damp cloth. Thoroughly dry surface
after cleaning.
Do not use fl ammable cleaners on Cart surfaces due to close proximity of electrical power and equipment.
All paints and plastic Cart components will withstand cleaning by most commonly used, diluted, non-abrasive solutions such as quaternary ammonia compounds, ammonia
enzyme cleaners, bleach or alcohol solutions.
• Pen and permanent and dry erase markers can be removed with 91% isopropyl alcohol and a soft cloth.
• Iodine stains can be removed with commonly used cleaners and a soft cloth.
• Never use steel wool or other abrasive materials that will damage the surface fi nish.
• Do not use strong solvents such as trichloroethylene and acetone. These solvents will damage the surface fi nish.
It is recommended that any cleaning solution be tested on a small, inconspicuous area to ensure surface is not harmed.
Adjustment, Service, Replacement - DO NOT attempt to adjust, service or replace any part of the StyleView Cart unless directed to do so through Ergotron-approved
documentation (i.e. installation instructions). Only Ergotron, Inc. or an Ergotron-certifi ed entity may adjust, service or replace StyleView Cart components. If any component on the
Cart is missing or damaged, the Cart must not be used, contact Ergotron Customer Care immediately to request a replacement part.
Cables - Keep cables neatly organized on the Cart (a variety of solutions are provided with your cart for this purpose). Excess cables should be routed away from moving
components with cable clips. Review Cable Routing section of this guide, or contact Ergotron Customer Care for more information.
Casters - Check casters periodically to make sure they are clean and free of debris that would prevent smooth travel. Avoid moving Cart across uneven, dirty or damaged
surfaces.
Customer Equipment- Make sure equipment is balanced and mounted securely to Cart. Do not reposition Cart components on riser or tower unless instructed to do so in the
installation instructions. Moving Cart components too high or too low on the Riser may create an unstable condition, leading to equipment damage or even personal injury. Contact
Ergotron Customer Care for information about moving Cart components.
Earth Bond Test: To ensure safety grounding between the
power cord ground connection and any accessible metal
parts on the cart, although not mandatory, the following test
is recommended to be performed every two years using a
calibrated medical device safety analyzer. The procedure is as
follows:
1. Disconnect the cart from mains power (unplug cart from the
wall).
2. Remove power from internal power system outlets by
pressing the power button on the Power System Interface.
3. Set up Earth Bond Test per the medical device safety analyzer
instructions.
4. Connect the Cart power cord to the medical device safety
analyzer.
5. Remove cover to access Lift Tension Adjustment point.
6. Attach second medical safety device analyzer probe to Lift
Tension Adjustment point on Cart.
7. Perform Test (25 amps).
8. Ensure Earth Bond is less than or equal to 0.2-ohms.
9. Remove medical device safety analyzer connections to Cart,
replace cover over Lift Tension Adjustment point, return the
Cart to service.
Insulation Resistance Test: To ensure mains power lines are
adequately insulated from earth ground, the following test,
although not mandatory, is recommended to be performed
annually using a calibrated medical device safety analyzer. The
procedure is as follows:
1. Disconnect the cart from mains power (unplug cart from the
wall).
2. Remove power from internal power system outlets by
pressing the power button on the Power System Interface.
3. Set up the Insulation Resistance Test per the medical device
safety analyzer instructions.
4. Connect the Cart power cord to the medical device safety
analyzer.
5. Perform test (500 V).
6. Ensure Insulation Resistance is greater than or equal to 1
Mohms.
7. Remove medical device safety analyzer connections to Cart
and return Cart to service.
Recommended Periodic Inspection and Maintenance
Component
Action
How often
By whom
ALD Batteries
(SV31 only)
Replace 4 ea AA NiMH batteries
1
Replace when
keypad #4 fl ashing
Any user
Ready On Batteries
(SV31 only)
Replace 2 ea AA Lithium batteries
1
Replace when light
on Ready On fl ashing
Any user
UI, Ethernet, USB
cables
Inspect for wear, pinching, bad
connectors
Monthly
Any user
Fan on side of CPU
compartment
Inspect for dust at intake, vacuum
as required using a vacuum cleaner
that DOES NOT generate ESD
(Electrostatic Discharge)
Monthly
Any user
Casters
Inspect for wear and debris
Monthly
Any user
Maximum Load
Inspect to ensure that maximum
recommended loads are not
exceeded
Daily
Any user
Power System SLA
batteries
Replace lead acid batteries
1,2
When instructed
by StyleLink or
when battery
runtime is 80% of
original runtime
IT Personnel
Power Module
Inspect for dust at intake, vacuum
as required using a vacuum cleaner
that DOES NOT generate ESD
(Electrostatic Discharge)
Monthly
IT Personnel
Battery harness
Inspect for wear, cracking, pinching,
or other damage
Monthly
IT Personnel
Coiled Cord
Inspect for wear, damage, or
stretching.
Weekly
Any user
1. Please dispose of all batteries in accordance with local law
2. Always replace with similar size battery (33 A-h to replace 33 A-h) and always replace in pairs