5
Safety notes
5.1
Precautions during Operation
Before using the unit, make sure that the Medical Product Consultant has conducted an introduction in
regard to function and safety precautions.
The unit should not be used if there are any doubts as to its being isolated from earth or the suitability
of the power cable.
The supplied power cable meets valid regulations for medical applications.
The unit is not intended for use in wet rooms, outdoors or in areas where there is danger of explosion.
Before the initial operation, the device must be adjusted using the base adjustors at the back to grant
absolute stability.
When the saddle is exchanged, make sure that the screws at the saddle are tightened hard enough so
that the saddle cannot be moved on the saddle pipe.
To move the handle bars and saddle, loosen the clamps and retighten them well afterwards. It is rec-
ommended to position the clamps with the levers positioned downwards. Secure clamping is ensured
when the handles are then again turned to this position during the adjustment procedure after every
movement.
The holding straps on the pedals must fit perfectly across the upper side of the shoe and be fastened
with a Velcro strap.
5.2
Safety Precautions when Operating with Other Devices
When several devices are coupled, there is a risk that the leakage currents may add up.
The RS-232 interface, which can be used for communication with other devices, is isolated to ensure
the patient’s safety.
External devices may only be connected with the interface cables supplied by ergosana.
Portable communication devices, HF radios and devices labelled with the symbol (non-ionic
electromagnetic radiation) can affect the operation of this device (see section 7.6).
5.3
Maintenance Safety Precautions
The device must be turned off and the power plug disconnected before cleaning with liquid cleaning
agents.
Only use standard cleaning agents for plastic surfaces.
The unit may only be opened, repaired and serviced by authorised and trained personnel.
5.4
Interference
The unit meets EMC regulations for medical products to ensure protection against emission and radia-
tion. Special caution should be taken when using this unit in combination with high-frequency devices.