- 44 -
↑
↓
RPM
Here you determine the RPM limits. When these limits
are exceeded, the LEDs for high or low speed (RPM) will
illuminate.
Select the value to change (Min. or Max.) and confi rm with
SELECT.
Using the arrow keys, change the corresponding value and
save the new value with SELECT.
S
ETTING
THE
RPM
LIMIT
VALUES
Note
•
The limits selected in this menu only apply to the load
range between 6 and 150 watts. At higher loads the RPM
limits automatically adapt to the respective loads:
Load (watts)
Green RPM range (1/min)
6 - 150
54 - 64 (adjustable)
151 - 250
58 - 65
251 - 350
68 - 75
351 - 450
78 - 85
451 - 550
88 - 95
551 - 650
98 - 105
651 - 750
108 - 115
751 - 850
118 - 125
851 - 950
> 125
951 - 999
> 130
RPM
Min
↑
0 ... 70
54
Max
↓
50 ... 130
64
Select
Choose function
P
ULSE
D
ISPLAY
The pulse readout on the display can be turned off .