- 23 -
+ 5 W
- 5 W
+ 5 W
- 5 W
PC Mode
Ergometry
Manual
Settings
Select
M
AIN
MENU
M
ANUAL
Use the softkeys on the right and left (
↑
↓
) to position the
bar cursor on MANUAL and confi rm the selection with
SELECT.
In this operating mode the user controls the entire exercise
test by selecting the loads, stage times and by initiating
blood pressure measurements.
The exercise test is started with the "Start" key, afterwards
the load can be set and changed with the +5 W and -5 W
keys (in increments of +/-1 W up to +/-25 W, as confi gured).
Blood pressure measurements an be initiated with
.
T
ERMINATING
AN
E
XERCISE
T
EST
The exercise test can be terminated manually at any time
with the END key located in the middle.
The load will immediately drop to 0 watt.
There is no recovery phase in the manual mode.
Start
I
NITIAL
SCREEN
OF
A
MANUAL
EXERCISE
TEST
S
CREEN
DISPLAY
DURING
THE
TEST
+ 5 W
- 5 W
End