Press the yellow pedal to activate cutting mode. Press the blue
pedal to activate coagulation mode. Make sure that the electro-
surgical unit responds correctly.
7
Cleaning, disinfection, sterilization
7.1 Safety
instructions
Erbe Elektromedizin recommends machine cleaning/
disinfection. Because it is much less effective, manual cleaning
is not recommended.
Not suitable for cleaning using an ultrasonic bath.
Never use sharp objects for cleaning.
Disinfectants must be rinsed off thoroughly after use.
Maximum air pressure during drying: 2 bar.
Maximum water pressure during rinsing: 2 bar.
During machine cleaning and disinfection, do not exceed 95 °C.
Observe the specifications from the manufacturer of the deter-
gent and disinfectant and in the chapter “Validated procedures
in overview“.
During sterilization, do not exceed 138 °C.
Do not sterilize in hot air.
7.2 Reprocessing
limitation
Frequent reprocessing can damage this product. Observe the
safety instructions for checking the product. With visible damage
or functional impairment, the product may no longer be used.
OBJ_DOKU-26720-006.fm Seite 20 Freitag, 7. April 2017 12:57 12
Содержание 20190-008
Страница 2: ...OBJ_DOKU 26720 006 fm Seite 2 Freitag 7 April 2017 12 57 12 ...
Страница 26: ...26 OBJ_DOKU 26720 006 fm Seite 26 Freitag 7 April 2017 12 57 12 ...
Страница 88: ...88 OBJ_DOKU 26720 006 fm Seite 88 Freitag 7 April 2017 12 57 12 ...
Страница 149: ...CS POKYNY K POUŽITÍ Epilační držáky 20190 008 20190 064 OBJ_DOKU 26720 006 fm Seite 149 Freitag 7 April 2017 12 57 12 ...
Страница 207: ...OBJ_DOKU 26720 006 fm Seite 207 Freitag 7 April 2017 12 57 12 ...