Steam sterilization
•
Fractionated vacuum method with adequate product drying
•
Holding time 3 to 18 minutes at 132 to 138 °C
•
Sterilizers in accordance with applicable national standards
and regulations (e.g. DIN EN 13060 or DIN EN 285)
•
Sterilization process validated according to DIN EN
ISO 17665
Please follow the recommendations issued by the sterilizer man-
ufacturer concerning loading, handling, and drying times.
7.11 Validated procedures in overview
The following procedures have been validated as suitable for
processing the product:
•
Manual cleaning/disinfection with the detergent Cidezyme
and disinfectant Cidex OPA (Johnson & Johnson Medical
Limited, Gargrave, Skipton).
– Cidezyme: Dosage 1.6%, 5 minutes exposure time at
room temperature.
– Cidex OPA: Dosage 0.3%, 12 minutes exposure time at
20 °C.
•
Machine cleaning and disinfection in a G 7836 CD disinfector
(Miele & Cie. KG, Gütersloh, Germany) with the cleaning
agent neodisher mediclean forte (Dr. Weigert GmbH & Co.
KG, Hamburg, Germany).
– neodisher mediclean forte: Dosage 0.5%, 5 minutes ex-
posure time at 55 °C.
– Disinfection: 5 minutes at 90 °C.
•
Sterilization in an autoclave Systec V-150 (Systec GmbH,
Linden/Germany).
– Steam sterilization with saturated steam, fractionated
vacuum procedure, 3 minutes, 132 °C, 3 bar.
Erbe recommends the processing procedures described in these
instructions for use. Equivalent different procedures are possible
OBJ_DOKU-26720-006.fm Seite 24 Freitag, 7. April 2017 12:57 12
Содержание 20190-008
Страница 2: ...OBJ_DOKU 26720 006 fm Seite 2 Freitag 7 April 2017 12 57 12 ...
Страница 26: ...26 OBJ_DOKU 26720 006 fm Seite 26 Freitag 7 April 2017 12 57 12 ...
Страница 88: ...88 OBJ_DOKU 26720 006 fm Seite 88 Freitag 7 April 2017 12 57 12 ...
Страница 149: ...CS POKYNY K POUŽITÍ Epilační držáky 20190 008 20190 064 OBJ_DOKU 26720 006 fm Seite 149 Freitag 7 April 2017 12 57 12 ...
Страница 207: ...OBJ_DOKU 26720 006 fm Seite 207 Freitag 7 April 2017 12 57 12 ...