Page 65
EPS Stromversorgung GmbH
Alter Postweg 101 • 86159 Augsburg
Germany
Fon: +49 821 / 570451-0
Fax: +49 821 / 570451-25
www.eps-germany.de
PSI 10000 4U Serie
Short Long
Description
Indication
OT
O
ver
Tem-
perature
Triggers an alarm if the internal temperature reaches a certain limit. The DC
output will be switched off which could be only temporarily, depending on the
setting
State after OT alarm
(see
Display,
analog &
digital inter-
face
MSP
M
aster-
S
lave
P
rotection
Triggers an alarm if the master unit loses contact to any slave unit. The DC
output will be switched off. The alarm can be cleared by reinitializing the MS
system or deactivating master-slave mode.
Display, dig-
ital interface
Safety
OVP
Safety
O
ver
V
oltage
P
rotection
Only featured with the 60 V model:
Triggers a special OVP alarm when the voltage on the DC output exceeds
the rigid threshold of 101% rated voltage. The DC output will be switched off.
For details refer to section
Display,
analog &
digital inter-
faces
SF
S
hare Bus
F
ail
Can occur in situations where the Share bus signal is damped too much
due to wrong or damaged (short-circuit) BNC cables or simply when at least
one of the Share bus connectors is wired to another device while the device
reporting the isn’t (yet) configured for master-slave operation. For details
Display, dig-
ital interfac-
es
► How to configure the thresholds of the adjustable device alarms
1.
While the DC output is switched off tap the touch area
on the main screen.
2.
In the menu tap on group
Protection
. On the right-hand side of the screen it will list all device alarms with
their adjustable thresholds. These are permanently compared to the actual values of voltage, current and
power on the DC output.
3.
Set the threshold for the protections relevant to your application if the default value of 110% is unsuitable.
The user also has the possibility of selecting whether an additional acoustic signal will be sounded if an alarm or
user defined event occurs.
► How to configure the alarm sound
(also see
“3.4.3. Configuration via the menu“
)
1.
Swipe with your finger up from the bottom edge of the screen or directly tap on the bottom bar:
2.
The quick menu will open. Tap on
to activate the alarm sound, or on
to deactivate it.
3.
Leave the quick menu.
3.6.2.1
User defined events
The monitoring functions of the device can be configured for user defined events. By default, events are deac
-
tivated (
Action
=
None
). Contrary to device alarms, the events only work while the DC output is switched on. It
means, for instance, that you can’t detect undervoltage (UVD) anymore after switching the DC output off and the
voltage is still sinking.
The following events can be configured independently:
Event
Meaning
Description
Range
UVD
U
nder
V
oltage
D
etection
Triggers an event if the output voltage falls below the
defined threshold.
0 V...U
Nom
OVD
O
ver
V
oltage
D
etection
Triggers an event if the output voltage exceeds the de
-
fined threshold.
0 V...U
Nom
UCD
U
nder
C
urrent
D
etection
Triggers an event if the output current falls below the
defined threshold.
0 A...I
Nom
OCD
O
ver
C
urrent
D
etection
Triggers an event if the output current exceeds the de
-
fined threshold.
0 A...I
Nom
OPD
O
ver
P
ower
D
etection
Triggers an event if the output power exceeds the de
-
fined threshold.
0 W...P
Nom