3.1.2.
STARTING UP THE UNIT
Starting up the unit consists in carrying out, in the following order, the INSTALLATION /
COMMISSIONING procedure described here-below.
1/ Press the ON/OFF push button
located on the front panel. The analyzer starts running, it passes
into warm-up cycle (duration of which is 1800 seconds maximum). The homepage is displayed: it allows
to visualize the warm-up progress period.
The warm-up period is ended when all the metrology parameters are within the operational limits. The
warm-up progress icon will disappear and the following homepage is displayed:
This screen is the analyzer standard function homepage. It contains additional pages that are displayed
by activating the buttons
to
, present at the top of the screen. These pages allow the
measurement channel display, the
« Real-time graph »
of the instantaneous measurement channel
values, and the
« Analyzer synoptic »
giving the main parameter values.
By using the touch screen mounted on the analyzer front panel: no password is required to activate the
standard functions.
By using a remote PC: a password is required to be input in the pop-up that will appear. Enter the User
password given in the «
Warning
» section of the Quick-Start, and validate with
.
Содержание AF22e
Страница 20: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 1 12 Page intentionally left blank...
Страница 22: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 2 2 Page intentionally left blank...
Страница 25: ...Duplication prohibited AF22e EnvironnementS A APRIL 2016 2 5 Figure 2 1 Diagram showing molecule energy levels...
Страница 76: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 3 42...
Страница 82: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 4 2 Page intentionally left blank...
Страница 98: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 4 18 Page intentionally left blank...
Страница 100: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 5 2 Page intentionally left blank...
Страница 107: ...Duplication prohibited AF22e EnvironnementS A APRIL 2016 5 9 Page intentionally left blank...
Страница 109: ...Duplication prohibited AF22e EnvironnementS A JUNE 2016 5 11 Figure 5 1 Interconnection board diagram...
Страница 113: ...Duplication prohibited AF22e EnvironnementS A JUNE 2016 5 15 Figure 5 4 Measurement board diagram...
Страница 118: ...EnvironnementS A AF22e Duplication prohibited JUNE 2016 5 20 Page intentionally left blank...