Duplication prohibited
AF22e
Environnement
S.A
APRIL 2016
5–7
Alarm message
(Min./Max. values)
Possible cause(s)
Possible corrective action(s)
Internal temperature
Min. value: 5°C
Max. value: 50°C
–
Ambient temperature is not within
the recommended range.
–
Install the analyzer in a 5-45°C
thermo-controlled room.
I UV lamp
Min. value: 45 mA
Max. value: 49 mA
–
UV lamp cable is disconnected.
–
Measurement board / UV lamp
power board cable is disconnected.
–
UV lamp power board is faulty.
–
Reconnect UV lamp cable.
–
Reconnect measurement board / UV
lamp power board cable.
–
Replace UV lamp power board.
HV PM
Min. value: 2400 mA
Max. value: 3200 mA
–
PM preamplifier / AF measurement
board cable is disconnected.
–
PM preamplifier board is faulty.
–
Connect PM preamplifier board /
measurement board cable.
–
Replace PM preamplifier board.
UV ref.
Min. value: 130 mV
Max. value:
–
UV lamp is not aligned.
–
UV lamp is faulty.
–
Aligned UV lamp as described in
chapter 4.
–
Replace UV lamp as described in
chapter 4.
NO Controller board
Min. value: absent
Max. value: N.A.
–
Controller board is unplugged.
–
Controller board is not compliant
with the analyzer, or firmware is
missing.
–
Controller board is faulty
–
Turn off the analyzer and disconnect
main cable. Reconnect controller
board ribbon.
–
Update controller board.
–
Replace controller board.
NO Measure board
Min. value: absent
Max. value: N.A.
–
Measurement board is
disconnected.
–
Measurement board is not compliant
with the analyzer, or firmware is
missing.
–
Measurement board is faulty.
–
Turn off analyzer and disconnect main
cable. Reconnect measurement board
ribbon.
–
Update measurement board.
–
Replace measurement board.
Содержание AF22e
Страница 20: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 1 12 Page intentionally left blank...
Страница 22: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 2 2 Page intentionally left blank...
Страница 25: ...Duplication prohibited AF22e EnvironnementS A APRIL 2016 2 5 Figure 2 1 Diagram showing molecule energy levels...
Страница 76: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 3 42...
Страница 82: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 4 2 Page intentionally left blank...
Страница 98: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 4 18 Page intentionally left blank...
Страница 100: ...EnvironnementS A AF22e Duplication prohibited APRIL 2016 5 2 Page intentionally left blank...
Страница 107: ...Duplication prohibited AF22e EnvironnementS A APRIL 2016 5 9 Page intentionally left blank...
Страница 109: ...Duplication prohibited AF22e EnvironnementS A JUNE 2016 5 11 Figure 5 1 Interconnection board diagram...
Страница 113: ...Duplication prohibited AF22e EnvironnementS A JUNE 2016 5 15 Figure 5 4 Measurement board diagram...
Страница 118: ...EnvironnementS A AF22e Duplication prohibited JUNE 2016 5 20 Page intentionally left blank...