65
1600751-4D
11.2.10.g
Shift
Shift is a special tool only available in the Myomed 6- series. The function of this parameter is to
zoom-in on a specific area in the graphic to have a better feedback for the patient. This is useful
for patient with muscle Hypertonia.
The adjustment is as follow:
•
Define the ideal bandwidth in the sensitivity setting screen.
•
Set the sensitivity back to the ideal bandwidth, the signal will disappear in the graphic.
Go back to the setting menu and select Shift.
•
Scroll with the shift until the signal is back on the screen. The sensitivity range is much
lower and only for a limited range.
Note!
•
The shift range is
linked to the Sensitivity
setting.
11.2.10.h
Myofe Stimulation
1. Place the reference electrode on the skin and connect it with the unit.
2. Place the EMG electrodes on the skin and connect these with the unit.
!
CAUTION
•
Connection of accessories other than the ones specified by the manufacturer can
adversely affect the safety of the patient and correct functioning of the equipment, and is
therefore not permitted.
Myofeedback
4
Continuous - Shift
0:00
0:00
E1
01:05
0.0
mA
0.0
mA
305
µV
Содержание 1600905
Страница 1: ...6 series User Manual ...
Страница 3: ...6 series User Manual ...
Страница 4: ......
Страница 5: ...3 1600751 4D ...
Страница 90: ...88 1600751 4D ...
Страница 92: ...90 1600751 4D 13 6 3 LP 13 6 4 CP 13 6 5 CPid ...
Страница 94: ...92 1600751 4D 13 7 4 IG50 Biphasic Envelope 13 7 5 IG100 Monophasic Envelope 13 7 6 IG100 Biphasic Envelope ...
Страница 95: ...93 1600751 4D 13 7 7 IG150 Monophasic Envelope 13 7 8 IG150 Biphasic Envelope ...