39
1600751-4D
10.3.2 Applying the fixation
Press the fixation included across the treatment head (see photograph hereunder)
Make sure that the ridge on the inner side of the fixation [A] falls in the groove of the StatUS
applicator / ultrasound treatment head [B]
10.3.3 Applying the gel pad (gel cushion)
Place a gel pad in the ring (see photograph). The StatUS applicator / ultrasound treatment head
is now ready for use.
The gel pads are intended for once-only use. Instructions are supplied with the gel pads. Read
these carefully.
You are also referred to chapter 10.3 for instructions on operating the device
.
Warning:
only use the original Enraf-Nonius gel pads! Do not use liquid gel as this will be
sucked into the vacuum tubes and pump, causing malfunction!
Any warranty claim will be rejected if liquid gel residue if found in any of these components.
Warning:
since the StatUS applicator / ultrasound treatment head transfers ultrasound energy
into the tissue at a most efficient and effective way, we advise you to use lower intensity settings
than the usual intensity settings for 'traditional' (semi-stationary or dynamic) ultrasound.