![Enraf Nonius 1600905 Скачать руководство пользователя страница 50](http://html.mh-extra.com/html/enraf-nonius/1600905/1600905_user-manual_2414071050.webp)
48
1600751-4D
11.2.5.a.1 Rheobase/ Chronaxie
•
Select Rheobase /
Chronaxie with the
central controller [5]
and enter with the
confirmation button [6]
•
Start with the
Rheobase. The phase
duration is fixed on
500ms.
•
To start push the
confirmation button [6]
you activate the
intensity setting. Use
the central controller [5]
to adjust the intensity.
•
Confirm the value by
pushing the
confirmation button [6]
•
Repeat by pushing the
confirmation button [6]
again.
•
Select Chronaxie with
the central controller [5]
and confirm with the
confirmation button [6]
•
The retrieved intensity
in the Rheobase will be
doubled automatic.
•
Adjust the Phase
duration with the
central controller [5]
until the muscle
contract.
Electrotherapy
2
Diagnostics
Electrode selection
Phase Interval
5 s
Rheobase / Chronaxie
S/D Curve Rectangular
S/D Curve Triangular
Report
Beep
mA
Diagnostics
2
Rheobase / Chronaxie
Rheobase
500 ms
Chronaxie
5.0
mA
Diagnostics
2
Rheobase / Chronaxie
Rheobase
500 ms
Chronaxie
1000 µs
10.0
mA
Содержание 1600905
Страница 1: ...6 series User Manual ...
Страница 3: ...6 series User Manual ...
Страница 4: ......
Страница 5: ...3 1600751 4D ...
Страница 90: ...88 1600751 4D ...
Страница 92: ...90 1600751 4D 13 6 3 LP 13 6 4 CP 13 6 5 CPid ...
Страница 94: ...92 1600751 4D 13 7 4 IG50 Biphasic Envelope 13 7 5 IG100 Monophasic Envelope 13 7 6 IG100 Biphasic Envelope ...
Страница 95: ...93 1600751 4D 13 7 7 IG150 Monophasic Envelope 13 7 8 IG150 Biphasic Envelope ...