Page 13 of 72
EN109-1498751-47 IFU
5 Installation
Inspection
In case of damage from transport is noticed, contact your local distributor.
DO NOT USE the device!
Immediately upon unpacking the device, perform the following steps:
•
Verify the delivery documents to make sure that the delivery is complete.
•
Verify that the packaging contains all the items listed in the standard accessories list.
•
Check the external components and accessories for possible damage due to transport.
System without a Vacotron
•
Remove the 4-series device and any additional items ordered from the carton and inspect for
damage that may have occurred during shipment.
•
Place the device on a desk or EN-Car. Ensure that there is sufficient air flow below the device (do
not place the device on a table-cover).
•
If required, place the unit on the supplied inclination foot to improve display legibility.
System with a Vacotron
•
Remove the vacuum unit and any additional items ordered from the carton and inspect for
damage that may have occurred during shipment.
•
Place the vacuum unit on a desk or EN-Car. Ensure that there is sufficient air flow below the device
(do not place the device on a table-cover).
•
Remove the 4-series device and any additional items ordered from the carton and inspect for
damage that may have occurred during shipment.
•
Place the main device on top of the vacuum unit.
•
Carefully lift the main device at the front and insert flat cable [17] into connector [18].
Connection to mains supply
Use of any other cable other than the supplied cable is strictly PROHIBITED as it
affects patient safety and the proper function of the device.
Do not place the device in a location where the power cord could be tripped over or pulled
out during treatment!
Do not attempt to use the device if it is not properly grounded. Make certain that the device
is electrically grounded by connecting it only to a grounded electrical service receptacle
conformable with the applicable national and local electrical codes regarding medical
environments!
•
Insert the mains cable into socket [1] and connect it to a wall socket.
•
Set power line switch [1] to On (1)
•
Power LED indicator [5] is lit green indicating that the device is connected to the mains supply.
•
Turn on the device with push button [4]
•
The device will initialize and perform a self-test. This may take a while.
•
At the end of the self-test the device enters the Home menu and is ready for use.
Placing the optional battery
Do not interchange the black and red wires as this will damage your device!
The battery contains material that is noxious to the environment. Observe the local
regulations when disposing of the battery!
Due to the high current demand of ultrasound applications, we recommend to explicitly use
batteries supplied by Enraf-Nonius B.V. (part number 2501.016).