Page 65 of 72
EN109-1498751-47 IFU
Vacuum leak
There probably is a leak in the vacuum system. This error is usually preceded by a continuously
running pump trying to reach the set vacuum. To protect the system, the pump is automatically
stopped after a certain amount of time. Inspect the vacuum cables and electrodes, set the vacuum
back to zero and try again. If the failure persists, stop using the device and contact your supplier.
Ultrasound applicator error
The ultrasound applicator has reported an error. Disconnect the device, wait some time and reconnect
it. If the error persists, stop using the device and contact your supplier.
Insufficient DC supply
This problem can sometimes occur with the small ultrasound applicator when operating from the
battery. If possible, continue the treatment with reduced therapy levels or connect the apparatus to
the mains supply.
Device needs to cool down before you can proceed
The temperature of the device is too high to start treatment. Remove the device from any potential
heat sources (such as direct sunlight) and let the device cool down. Restart the device to continue.
Possible risk of skin irritation
This notification is shown when operating a waveform that contains a DC component. Operator
should proceed with caution.
The intensity you have set is above the recommended maximum
When increasing the dosage of a StatUS device above the recommended maximum of 1.5W/cm
2
this
notification is shown. Operator should proceed with caution.
StatUS applicator error
The StatUS applicator has reported an error. Disconnect the device, wait some time and reconnect it. If
the error persists, stop using the device and contact your supplier.
The modulation is OFF. Higher risk of cavitation and/or hotspots
This notification is shown when operating a StatUS with modulations turned off. Turning off the
modulation increases the risk of cavitation and / or hotspots. Operator should proceed with caution.
Battery empty. Use power cord
The battery has been completely drained and operation of the device is disabled. Please connect the
power cord and charge the battery. Restart the device when connected to mains power to continue
operation.
Combination Therapy is only allowed with one US head connected
The notification will be shown when there is more than 1 ultrasound head connected to the device.
Combination therapy is only allowed with a single ultrasound head connected. Disconnect the inactive
ultrasound head to continue.
Combination therapy is not allowed with StatUS
The notification will be shown when there is a StatUS applicator connected to the device. Combination
therapy is not allowed with a StatUS applicator connected. Disconnect the StatUS applicator to
continue.