Installation and Operation Manual - COMTEC
®
6000
Installation
Doc.-ID: COM6000_11022020_EN
9
2
Installation
Warning
The electronic unit does not have a line voltage main switch. The line voltage power supply requires a
switch or breaker. The line voltage switch/fuse/breaker must be in accordance with the local technical
standards and should be in close proximity to the electronic unit and must be identifiable as such.
The probe cable is suitable for an ambient temperature range from -40°C to +90°C. All other installed
cables must be suitable for the ambient temperature range at side and must have the required size. All
electronic unit terminals are specified from 0,08mm² (AWG 26) to 2,5mm² (AWG 14). If wire end ferrules
are used the next smaller size is required. Before removal of the electronic terminal cover the line voltage
must be switched off. The line voltage to the electronic unit must be switched on again after the cover is
back in position. After installation, power conducting parts may not be accessible.
2.1
Installation requirements for electronic unit
Keep the minimum distance to adjacent objects
Install at eye level
Avoid vibrations greater than 2g
Mind the IP code
Heavy equipment, ensure
proper lifting and carrying
Ambient temperatures
Min.: -20 °C (-4 °F) / Max.: +55 °C (+131 °F) - (Pump version 20°C - +50°C)
Drilled holes for the electronic unit
Use suitable screws
Electronic unit
Figure 4
- Installation of the Electronic unit