Status Messages
Doc.-ID: COM6000_11022020_EN
45
Error Messages
(cont.)
Error
Message
Relay
contact
O
2
signal
output
Description
O2 sensor
offset too
high
System
error,
open
The error can only occur during an O2 sensor calibration. An error reset can be
carried out by the user. A successful re-calibration also resets the error.
Possible causes:
defective false test gas (not applicable with test air), test gas
flow too low, O2 sensor defective.
O
2
sensor
slope too
low
System
error,
open
The error can only occur only during an O
2
sensor calibration. An error reset can
be carried out by the user. A successful re-calibration also resets the error.
Possible causes:
Incorrect calibration gas, test gas flow too low, probe filter
damaged, filter head missing, faulty O
2
sensor.
O
2
sensor
slope too
high
System
error,
open
The error can only occur only during an O
2
sensor calibration. An error reset can
be carried out by the user. A successful re-calibration also resets the error.
Possible causes:
Incorrect calibration gas, O
2
sensor defective.
O
2
sensor
signal
instable
System
error,
open
The error can only occur only during an O
2
sensor calibration. An error reset can
be carried out by the user. A successful re-calibration also resets the error.
Possible causes:
test gas flow too low, probe filter is damaged, process
pressure fluctuations too high.
CO
e
board
communicatio
n error
20.80mA
unless set
otherwise
The error can occur at any time and signals a failure in communication between
the display circuit board and the CO
e
circuit board. A reset can be carried out by
the user. If the error occurs again, switch the system on and off. If the error
persists after restarting the system, contact an ENOTEC service point.
CO
e
sensor
heater short
circuit
20.80mA
unless set
otherwise
The error can occur at any time and signals a short circuit either by CO
e
sensor
heating or within the connection between the electronic unit and CO
e
sensor. A
reset can be carried out by the user.
Possible causes:
Short-circuit of the CO
e
sensor heater cable at the terminal
points of the electronics or the sensor, short circuit within the probe cable, short
circuit inside the COe sensor
CO
e
sensor
heater short
circuit
20.80mA
unless set
otherwise
The error can occur at any time and signals a break in the circuit of CO
e
sensor
heater. A reset can be carried out by the user, once the fault has been remedied.
Possible causes:
contact problems between CO
e
sensor cable and terminal
points of the electronics respectively the probe, probe cable is damaged, defective
COe sensor
Open circuit
CO
e
sensor
heating
20.80mA
unless set
otherwise
The error can occur at any time and signals a break in the circuit of CO
e
sensor. A
reset can be carried out by the user, once the fault has been remedied.
Possible causes:
contact problems CO
e
sensor wire to the terminal points of the
electronics and the probe, probe cable is damaged, defective CO
e
sensor
Open circuit
CO
e
sensor
20.80mA
unless set
otherwise
The error can occur at any time and signals a failure of the electronic components
of the CO
e
board. A reset cannot be carried out by the user. If the error persists
after restarting the system, contact an ENOTEC service point.
COe board
hardware
error A-C
20.80mA
unless set
otherwise
CO
e
sensor calibration failed for one of the following reasons. A reset cannot be
carried out by the user. All entries triggered by this error will then also be deleted.
CO
e
sensor
calibration
failed
20.80mA
unless set
otherwise
CO
e
sensor calibration failed for one of the following reasons. A reset cannot be
carried out by the user. All entries triggered by this error will then also be deleted.
CO
e
sensor
offset too
low/high
20.80mA
unless set
otherwise
The error can only occur only during COe sensor calibration. A reset cannot be
carried out by the user. A successful re-calibration remedies the error. Possible
causes: insufficient supply of reference air.
CO
e
factor
alpha/beta/
gamma
invalid
20.80mA
unless set
otherwise
The error can only occur only during CO
e
sensor calibration. It can be reset by the
user. A reset cannot be carried out by the user. A successful re-calibration also
remedies the error.
Possible causes:
inadequate reference air supply, process pressure is too high,
incorrect test gas (not test air), CO
e
sensor defective.
Error
ENOTEC
REMOTE
module
System
error,
open
Indicates a hardware error of the ENOTEC REMOTE
module.
Possible cause:
the ENOTEC REMOTE
module is defective.
Unlisted errors: Other errors cannot be remedied by the customer.
Please contact an ENOTEC service point