3.1.1 Wiring the AM-5175 

The electrical installation should conform to appropriate electrical codes, such as the National Electrical Code in the United 
States. 

W

ARNING

:

 

The compliance of the installation to appropriate codes is not 

ENMET

’s responsibility. 

The 

AM-5175

 should be powered through circuit breakers provided for this purpose. 

3.1.2 Power Supply 

The input power can vary from 100 to 240 

V

AC

, 50/60 Hz. Power should be connected to the Power Input Terminal 

TB1

 and 

the Ground screw.  See Figure 4 for location.  Instrument is supplied with a strain relief and a standard U.S. 3 prong power line 
cord. 

For DC wiring 24

V

DC

 may be wired to J12, (J12-1)position 1 + with ground connected to (J12-2)position 2. 

Upon supplying power to the 

AM-5175

 

The green power on LED is lit. 

 

The display backlight is lit, and instrument will step through a start-up sequence: unit serial number and software revision 
may be shown on the display. 

The instrument may go into alarm briefly, but the sensors stabilize quickly.  If the instrument persists in alarm, acknowledge the 
alarm by pressing the 

S

ELECT 

button.  If alarm persists longer than 30 minutes, call 

ENMET

 customer service personnel. 

W

ARNING

:

 

Continuous gas detection and alarm systems (110

V

AC

/220

V

AC

 / 24

V

DC

/

12

V

DC

 powered) become inoperative upon 

loss of primary power.  Contact factory for specifications and pricing of backup battery systems. 

 
AC Power Supply Terminal: TB1 

 

Label on PCB 

Function 

110

V

AC

 

TB1 ACN 

Neutral 

 

TB1 ACL 

Line 

Ground 
Screw 

 

AC GND 

220

V

AC

 

TB1 ACN 

Neutral 

Optional 

TB1 ACL 

Line 

Ground 
Screw 

 

AC GND 

C

AUTION

: 110/220 

V

AC

 when applied 

 

DC Power Supply/4-20mA 

 

Terminal: J12 

Position 

Function 

1 + 

24 

V

DC

 

power 

GND  

4 - 20 mA out 

 

 

 

Sensor Terminal: J8 

Position 

Function 

Color 

1 + 

Sensor 

Red 

Sensor Signal 

White 

Sensor Ground 

Black 

 

Figure 4: Power Terminal Connections AM-5175 

Cover Inside View 

Opened Upward 

Caution

110/220 

Vac

 

when applied 

Содержание AM-5175

Страница 1: ...ENMET Corporation PO Box 979 Ann Arbor MI 48106 0979 www enmet com Manual part number 80003 551 MCN 14 003 03 20 14 AM 5175 Ambient Air Monitor Manual...

Страница 2: ...Sensor Replacement 15 5 3 1 A Factory calibration must be performed 15 6 0 Accessory and Replacement Parts 16 7 0 Technical Data and Specifications 16 8 0 WARRANTY 17 List of Tables Table 1 Relay Fai...

Страница 3: ...tems packing material and carton As Is Within 5 days of receipt notify the carrier s local office and request immediate inspection of the carton and the contents After the inspection and after you hav...

Страница 4: ...information is displayed Audio Alarm Horn Audio alarm 105 dB at 30cm 12in The audio alarm is activated when the unit is in alarm Visual Indicators and Alarms LED indicators Power Fault Indicator LED G...

Страница 5: ...14 J15 J16 J17 This group of terminals is located on the Circuit Board For the contacts for each of three alarm relays and for the contacts of a fault relay See Section 3 3 Terminal J12 For VDC back u...

Страница 6: ...riate vertical surface leaving room for lid to be opened using the mounting holes provided Avoid areas with excessive vibration or temperature extremes The holes in the bottom of the enclosure are 0 1...

Страница 7: ...through a start up sequence unit serial number and software revision may be shown on the display The instrument may go into alarm briefly but the sensors stabilize quickly If the instrument persists i...

Страница 8: ...ly Open J14 K1 Relay 1 NC Normally Closed J14 K1 Relay 1 COM Common Alarm 2 J15 K2 Relay 2 NO Normally Open J15 K2 Relay 2 NC Normally Closed J15 K2 Relay 2 COM Common Alarm 3 J16 K3 Relay 3 NO Normal...

Страница 9: ...tatus See the Section 4 1 1 below 4 1 1 Typical Start Up When power is supplied to the AM 5175 the instrument will display the following sequence of information Typical start up sequence of informatio...

Страница 10: ...Display indicates mA Span range Full Scale Press MENU button Display returns to gas measurent Operational Display Flow Chart 4 2 1 Alarm Conditions AM 5175 There are three alarm set points available T...

Страница 11: ...AM 5175 Maintenance Menus Sequence Example of Display Function Normal Display Mode Measurement of CO Press and hold the MENU button for 2 4 seconds to enter the Maintenance Menu The Power Fault LED wi...

Страница 12: ...ess the Menu button to cycle through Maintenance Menus Select OR Zero Cal OK If the Zero signal is not within Preset Specs the AM 5175 will display Bad ZERO Menu Bad ZERO If the Zero signal is within...

Страница 13: ...is recommended that the Zero procedure be done prior to the Span procedure ENMET Corporation recommends at least quarterly calibration of the AM 5175 instrument Calibration equipment is available fro...

Страница 14: ...in 1 2 seconds display will change to Span If you wish to Span the sensor press the SELECT button you are now ready to apply gas Proceed to gas span step 2 If you wish to Exit the maintenance menu pre...

Страница 15: ...To exit cal press MENU button until Exit appears and press SELECT button If the sensor is outside of acceptable parameters Bad Span is displayed If the sensor did not respond an incompatible span gas...

Страница 16: ...intenance menu as shown in Figure 6 AM 5175 Maintenance Menu flow chart 1 Press the MENU button until to display Span is displayed 2 Press the SELECT button to initiate the mA Span menu 3 Press the ME...

Страница 17: ...is present remove the clip before plugging in the new sensor 6 Replace the Sensor PC board and 2 retaining screws 7 Reconnect the instrument to power 8 Recalibrate the instrument See Section 5 2 Step...

Страница 18: ...num Cylinders 67028 0200 Sensor H2S 67028 1100 Sensor O2 67028 1200 Sensor CO 73083 000 Carry Case 7 0 Technical Data and Specifications Electrical Power 15 Amp fused branch circuit 100 240 VAC 0 45A...

Страница 19: ...sed by the improper use of the product The purchaser indemnifies and saves harmless the company with respect to any loss or damages that may arise through the use by the purchaser or others of this eq...

Страница 20: ...processing delays ENMET charges a one hour minimum billing for all approved repairs with additional time billed to the closest tenth of an hour All instruments sent to ENMET are subject to a minimum...

Страница 21: ...nt Terms K COD Check one K VISA MasterCard______________________ ________ ________ Card number Expiration Card Code K American Express______________________ ________ ________ Card number Expiration Ca...

Отзывы: