5.2.3 Gas Span 

It is recommended that the Zero Function be performed first. 

Do not perform a calibration unless span gas is applied to sensor.  Calibration can be aborted by pressing and holding the 

M

ENU

 

button for 3 – 4 seconds. 

Enter the maintenance menu.  See Figure 6, 

AM-5175

 Maintenance Menu flow chart. 

1. 

Press the 

M

ENU

 

button until Span display. 

2. 

Press the 

S

ELECT

 

button to perform a Span procedure.   

The display will alternate between the calibration gas concentration (Example

Cal 20

) and a signal level (

PV

). 

 To Abort calibration press and Hold 

M

ENU

 button for 3 – 4 seconds, “Abort?” will appear, press 

S

ELECT

 button to return 

to Span. 

 To change calibration gas level to be used, press and Hold 

S

ELECT

 button for 3 – 4 seconds, use menu button to change 

digit and select button to move to next digit. 

3. 

Attach the associated calibration gas cylinder to the calibration adapter.  See Figure 7 calibration adapter. 

4. 

Open the valve to apply the calibration gas to the sensor. 
An auto detect sequence is initiated after 30 seconds, the 

AM-5175

 will monitor the cal reading for stability. 

5. 

Watch for the signal level to stabilize.  This should take about 1 – 4 minutes. 

6. 

Once the signal level has stabilized, 

 

If the Span is successful, Cal OK appears momentarily, then will advance to Alarm1 menu. 
To exit cal, press 

M

ENU

 button until Exit appears and press 

S

ELECT

 button 

 

If the sensor is outside of acceptable parameters, Bad Span is displayed. 

 

If the sensor did not respond, an incompatible span gas was applied and the sensor did not respond at all,  
Same mV is displayed then will return to Span. 

 

If calibration is not successful, it is suggested that calibration be attempted again in 30-60 minutes. 
 

If the sensor will not calibrate See Section 5.3 for sensor replacement. 

7. 

Remove the calibration gas. 

8. 

Press the 

M

ENU

 button to advance to Exit menu or desired menu. 

N

OTE

:

 Software revision may cause variations of display output. 

Example of Calibration Display: 
 

Menu 

PV: 

 

20 

Apply Cal Gas until signal value becomes stable  (about 1 to 4 

minutes) See Figure 7 Calibration Adapter 

 

When cal signal is stable AM-5175 will automatically update: 
 

If cal is good display will indicate OK or Same and advance to 
Alarm1 

 

If cal is not within preset “range” display will indicate Bad Sens 
or Same mV The AM-5175 will return to the Span Menu 

To exit press Menu button until Exit appears and press Select 

Cal  OK 

Span 

PV: 

Same  mV 

Bad  Sens 

OR 

Select 

Select 

Note:

 To change calibration gas level.  Press and Hold the Select button to change the Calibration Gas Level 

-Use the Menu button to change digits 
-Use the Select button to move to next digit 

Select 

  20 

Note:

 To abort calibration.  Press and Hold the Menu button to abort Calibration 

When “Abort?” appears press Select button to advance to desired menu 

Abort? 

Select 

Menu 

Содержание AM-5175

Страница 1: ...ENMET Corporation PO Box 979 Ann Arbor MI 48106 0979 www enmet com Manual part number 80003 551 MCN 14 003 03 20 14 AM 5175 Ambient Air Monitor Manual...

Страница 2: ...Sensor Replacement 15 5 3 1 A Factory calibration must be performed 15 6 0 Accessory and Replacement Parts 16 7 0 Technical Data and Specifications 16 8 0 WARRANTY 17 List of Tables Table 1 Relay Fai...

Страница 3: ...tems packing material and carton As Is Within 5 days of receipt notify the carrier s local office and request immediate inspection of the carton and the contents After the inspection and after you hav...

Страница 4: ...information is displayed Audio Alarm Horn Audio alarm 105 dB at 30cm 12in The audio alarm is activated when the unit is in alarm Visual Indicators and Alarms LED indicators Power Fault Indicator LED G...

Страница 5: ...14 J15 J16 J17 This group of terminals is located on the Circuit Board For the contacts for each of three alarm relays and for the contacts of a fault relay See Section 3 3 Terminal J12 For VDC back u...

Страница 6: ...riate vertical surface leaving room for lid to be opened using the mounting holes provided Avoid areas with excessive vibration or temperature extremes The holes in the bottom of the enclosure are 0 1...

Страница 7: ...through a start up sequence unit serial number and software revision may be shown on the display The instrument may go into alarm briefly but the sensors stabilize quickly If the instrument persists i...

Страница 8: ...ly Open J14 K1 Relay 1 NC Normally Closed J14 K1 Relay 1 COM Common Alarm 2 J15 K2 Relay 2 NO Normally Open J15 K2 Relay 2 NC Normally Closed J15 K2 Relay 2 COM Common Alarm 3 J16 K3 Relay 3 NO Normal...

Страница 9: ...tatus See the Section 4 1 1 below 4 1 1 Typical Start Up When power is supplied to the AM 5175 the instrument will display the following sequence of information Typical start up sequence of informatio...

Страница 10: ...Display indicates mA Span range Full Scale Press MENU button Display returns to gas measurent Operational Display Flow Chart 4 2 1 Alarm Conditions AM 5175 There are three alarm set points available T...

Страница 11: ...AM 5175 Maintenance Menus Sequence Example of Display Function Normal Display Mode Measurement of CO Press and hold the MENU button for 2 4 seconds to enter the Maintenance Menu The Power Fault LED wi...

Страница 12: ...ess the Menu button to cycle through Maintenance Menus Select OR Zero Cal OK If the Zero signal is not within Preset Specs the AM 5175 will display Bad ZERO Menu Bad ZERO If the Zero signal is within...

Страница 13: ...is recommended that the Zero procedure be done prior to the Span procedure ENMET Corporation recommends at least quarterly calibration of the AM 5175 instrument Calibration equipment is available fro...

Страница 14: ...in 1 2 seconds display will change to Span If you wish to Span the sensor press the SELECT button you are now ready to apply gas Proceed to gas span step 2 If you wish to Exit the maintenance menu pre...

Страница 15: ...To exit cal press MENU button until Exit appears and press SELECT button If the sensor is outside of acceptable parameters Bad Span is displayed If the sensor did not respond an incompatible span gas...

Страница 16: ...intenance menu as shown in Figure 6 AM 5175 Maintenance Menu flow chart 1 Press the MENU button until to display Span is displayed 2 Press the SELECT button to initiate the mA Span menu 3 Press the ME...

Страница 17: ...is present remove the clip before plugging in the new sensor 6 Replace the Sensor PC board and 2 retaining screws 7 Reconnect the instrument to power 8 Recalibrate the instrument See Section 5 2 Step...

Страница 18: ...num Cylinders 67028 0200 Sensor H2S 67028 1100 Sensor O2 67028 1200 Sensor CO 73083 000 Carry Case 7 0 Technical Data and Specifications Electrical Power 15 Amp fused branch circuit 100 240 VAC 0 45A...

Страница 19: ...sed by the improper use of the product The purchaser indemnifies and saves harmless the company with respect to any loss or damages that may arise through the use by the purchaser or others of this eq...

Страница 20: ...processing delays ENMET charges a one hour minimum billing for all approved repairs with additional time billed to the closest tenth of an hour All instruments sent to ENMET are subject to a minimum...

Страница 21: ...nt Terms K COD Check one K VISA MasterCard______________________ ________ ________ Card number Expiration Card Code K American Express______________________ ________ ________ Card number Expiration Ca...

Отзывы: