GETTING FAMILIAR WITH THE ADS 2000
In order to operate the ADS 2000 properly, you need know its controls.
Front View:
LCD DISPLAY
- Displays Minute Volume, Inspiratory Time, Proximal Airway Pressure, Flow Rate,
Breaths Per Minute, Peak Inspiratory pressure, and Assist Pressure.
12
SET
POWER
switch - This switch turns the power to the ADS 2000
ON
and
OFF
.
MASK
switch - This switch is used for enabling and disabling the
Mask Mode
SET / RUN
Switch - When in the
SET,
this switch allows the operator to enter in
the patients' weight. When switched to
RUN
the ADS 2000 begins ventilating the
patient.
VOLUME
Knob - This knob controls the volume of the audible
warning alarm
. We suggest that you
start with it turned fully clockwise, i.e. full volume and adjust it as necessary.
FILL / HOLD
Button - When this button is pressed, the ADS 2000 will fill the lungs of the patient to
the indicated peak inspiratory pressure and hold it until the button is released.
BREATHE
Button - When this button is pressed, the ADS 2000 will initiate a breath to the
indicated peak inspiratory pressure to the patient.
WEIGHT UP / DOWN
Buttons - When the ADS 2000 is in the
SET MODE
these buttons allow the
operator to enter the weight of the patient, from below
1.0 Kg
. to
68 Kg
.
FLOW RATE UP / DOWN
Buttons - Depressing these buttons cause the
FLOW RATE
to increase
or decrease, from
0.2 LPM
to
60 LPM
.
BREATHS PER MINUTE UP / DOWN
Buttons - These buttons control the minimum number
of times that the ADS 2000 will breathe per minute, from 1 to 95 B.P.M.
P.I.P. UP / DOWN
Buttons - These buttons control the
Peak Inspiratory Pressure
in cm of
H
2
O, that the ADS 2000 will deliver to the patient, from
5.0 cm
to
35 cm
./ H
2
O
ASSIST UP / DOWN
Buttons - These buttons set the
sensitivity of inspiratory effort
necessary for the ADS 2000 to facilitate an assisted breath, from
–0.3 to 6.0 cm
. /
H
2
O They also allow the
ASSIST
feature to be turned off.