UNDERSTANDING
PEEP MODE
1. The ADS 2000 has a built in
PEEP
(
Positive End Expiratory
Pressure
) mode. To activate the
PEEP
MODE
perform the following:
a. Press the
Assist UP
button until the unit displays “
Entering the PEEP MODE
”
and the alarm beeps.
b. As shown below, the LCD display will now show a
PEEP
value instead of an
Assist
number
NOTE: An " * " (asterisk) is displayed as an indication that the ADS 2000 is in the
PEEP
MODE
.
NOTE: The
Assist MODE
will not function when the ADS 2000 is in
PEEP
MODE
.
NOTE:
PEEP
pressures range from 0 to 9 cm. of H
2
O. Adjust the
PEEP
pressure by using the
ASSIST UP
and
ASSIST DOWN
buttons.
c. To exit the
PEEP
MODE
Press the
ASSIST DOWN
button until the
unit displays “
Entering the ASSIST MODE
” and the alarm beeps.
20