Connecting the gas sampling system
- To connect the external gas sampling system
perform the following steps:
1) Insert the Luer lock connector to the gas sampling input and rotate it clockwise one half turn.
2) Connect the two breathing circuits ends to the breathing circuit ports.
3) Now insert the gas-sampling elbow into the end of the breathing circuit as shown.
4) Insert the Luer lock connector to the gas-sampling elbow and rotate it clockwise one half
turn.
Note: You’re ADS 2000 unit is fitted with a hose disconnect safety system. In the event that the
external sampling hose becomes disconnected, the system will disconnect the gas input to the
patient, sound an alarm and flash the red LED on the front left corner of the unit.
Once the external sampling hose is reconnected, the unit will continue to work at its previous setting.