55
THE ENERAC 700 COMMAND SET
DATA COMMANDS
COMMAND
FUNCTION
ATEM?
Returns the present value of ambient temperature.
BATT?
Returns the present battery voltage.
CDOX?
Returns the present value of carbon dioxide.
CMNX?
Returns the present value of carbon monoxide.
COMB?
Returns the present value of combustible gases.
DRAF?
Returns the present value of stack draft.
EFFI?
Returns the present value of combustion efficiency.
EXAR?
Returns the present value of excess air.
NOXY?
Returns the present value of nitric oxide (NO).
NO2Y?
Returns the present value of nitrogen dioxide (NO
2
).
NOXX?
Returns the present value of oxides of nitrogen (NO
X
).
OXYG?
Returns the present value of oxygen.
SO2X?
Returns the present value of sulfur dioxide.
STEM?
Returns the present value of the stack temperature
TEXT?
Returns a complete record of all current stack parameters.
VELO?
Returns the present value of stack velocity or flow rate.
SETUP COMMANDS
COMMAND
FUNCTION
ATOF?
Returns the ambient temperature offset in °C.
ATOF XX
Sets the ambient temperature offset to XX°C.
COOL?
Returns the thermoelectric cooler duty cycle.
COOL XX
Sets the thermoelectric cooler duty cycle:
XX=50 50% power
XX=100 100% power
CORF?
Returns the temperature units.
CORF X
Sets the temperature units:
X=F
Fahrenheit
X=C
Celsius
CUST?
Returns the customer name. This name appears on the display and
all printouts.
CUST XXXX
Sets the customer name, up to 21 characters long.
DATE?
Returns the present date.
DATE
XX/XX/XX
Sets the present date.
FUEL?
Returns the current fuel used.
FUEL NN?
Returns the fuel currently stored in location #NN.
FUEL NN
Changes its current fuel to fuel #NN (1-15).
MODE?
Returns the current emissions units.
Содержание 700
Страница 2: ...2 ENERAC 700 MANUAL EDITION 4 M700 FIRMWARE 3 0 M 700 HARDWARE H ...
Страница 4: ...4 APPENDIXC EPA TEST METHODS 70 ...
Страница 14: ...14 ...
Страница 18: ...18 ...
Страница 20: ...20 ...
Страница 49: ...49 ...
Страница 63: ...63 ...
Страница 67: ...67 ...