Commissioning
Teqwave H
48
Hauser
Parameter overview with brief description
Parameters
Procedure
Selection/input
Factory setting
Output mode
Select the mode for the switch point
definition.
• Range mode
• Trigger mode
Range mode
If
Range mode
is selected, enter the upper and lower limit to define the switch points.
H
M
NC
0
1
H
Min
Max
A0035471
17
Example: relay output configured in the range mode as an NC contact
NC Normally closed
0
Closed relay
1
Open relay
M
Measured value
Min Switch point Min
Max Switch point Max
H
Hysteresis
If
Trigger mode
is selected, enter the value to define the switch point.
H
M
NC
0
1
P
A0035472
18
Example: relay output configured in the trigger mode as an NC contact
NC Normally closed
0
Closed relay
1
Open relay
M
Measured value
P
Switch point
H
Hysteresis
Switch point
Prerequisite:
Trigger mode
is selected in the
Output mode
parameter.
Set the value at which the relay
changes its switch status, while
paying attention to the measuring
range limits → 80 and the
concentration app data sheet.
Positive decimal
• Analysis parameter 1…n:
depends on the selected
concentration app
• Temperature: 60.00
• Speed of sound: 1500.00
• Density: 1000.00
Switch point Max
Prerequisite:
Range mode
is selected in the
Output mode
parameter.
Set the upper limit at which the
relay changes its switch status,
while paying attention to the
measuring range limits → 80
and the concentration app data
sheet.
Positive decimal
• Analysis parameter 1…n:
depends on the selected
concentration app
• Temperature: 120.00
• Speed of sound: 2000.00
• Density: 1500.00