System integration
Teqwave H
40
Hauser
Register name
Access Data type
Register address Length Input/output
Concentration C
Read
IEEE754 32 bit
0x0004:0x0005
2
Additional analysis
parameter 1
Read
IEEE754 32 bit
0x0006:0x0007
2
Additional analysis
parameter 2
Read
IEEE754 32 bit
0x0008:0x0009
2
Additional analysis
parameter 3
Read
IEEE754 32 bit
0x000A:0x000B
2
Additional analysis
parameter 4
Read
IEEE754 32 bit
0x000C:0x000D
2
Additional analysis
parameter 5
Read
IEEE754 32 bit
0x000E:0x000F
2
Temperature / K
Read
IEEE754 32 bit
0x0010:0x0011
2
Speed of sound / m/s
Read
IEEE754 32 bit
0x012:0x0013X
2
Density / kg/m³
Read
IEEE754 32 bit
0x0014:0x0015
2
Dispersion
Read
IEEE754 32 bit
0x001A:0x001B
2
System status
Read
32 bit unsigned
0x0050:0x0051
2
Diagnostic
information
To increase measuring accuracy, predefined compensation values can be used for the compensation of the
speed of sound and acoustic density specifically for your application. The compensation values can be read
into the device from external measuring devices via Modbus TCP or entered as constant measured values
(Configuration of compensation value 1…4). The compensation values are only available if they were
provided in the data record of the concentration app upon delivery.
Compensation value 1
Read /
write
IEEE754 32 bit
0x1852:0x1853
2
Floating point
number in IEEE754
format (big endian)
Compensation value 2
Read /
write
IEEE754 32 bit
0x1854:0x1855
2
Compensation value 3
Read /
write
IEEE754 32 bit
0x1856:0x1857
2
Compensation value 4
Read /
write
IEEE754 32 bit
0x1858:0x1859
2
Configuration of
compensation value 1
Read /
write
16 bit unsigned
0x1D6B
1
• 0 = disabled:
measured
variable is not
taken into
consideration.
• 1 = fixed value:
entered
measured value
is taken into
consideration.
• 2 = Modbus: the
measured value
received via
Modbus TCP is
taken into
consideration.
Configuration of
compensation value 2
Read /
write
16 bit unsigned
0x1D86
1
Configuration of
compensation value 3
Read /
write
16 bit unsigned
0x1DA1
1
Configuration of
compensation value 4
Read /
write
16 bit unsigned
0x1DBC
1
While values are transmitted in "big endian" format during Modbus communication, in
line with specifications, the sequence of the bytes received may have to be reversed,
depending on the system used, to convert between big endian and little endian
format.
Example: read concentration A
The result is a floating point number in the format IEEE754 32 bit. The value is saved in
two consecutive addresses, with the first address containing the least significant word
Содержание Teqwave H