Technical data
Teqwave H
80
Hauser
16
Technical data
16.1 Application
The measuring device is intended only for the measurement or analysis of liquids.
To ensure that the device remains in proper operating condition for its service life, use the
measuring device only for media against which the process-wetted materials are
sufficiently resistant.
16.2 Function and system design
Measuring principle
Concentration measurement using ultrasonic waves.
Measuring system
For information on the structure of the device, see "Product description" → 11.
16.3 Input
Measured variables
Measured variables
• Speed of sound
• Temperature
• Dispersion
• Acoustic density
Analysis parameters available for order
• Concentration
• Concentration ratio (ratio between two concentrations)
• Calculated consistency
Measuring ranges
Speed of sound
600 to 2
000 m/s
Temperature
0 to 120 °C (32 to 248 °F)
Acoustic density
0.7 to 1.5 g/cm³
Analysis parameter
As per the concentration app data sheet
Input signal
Digital input
Function
Select analog channel 1 to 4;
The digital inputs "0" and "1" can only be connected to the signal ground.
Version
Open and ground
Do not connect external voltage to these terminals.