24.
Reset the scrubber lifetime monitor with the
New Scrub Installed
parameter and the
General Fault Alarm
with the Reset option for
the
General Alarm
DO
parameter (refer to the Firmware Manual).
25.
Re-validate the system with an appropriate gas standard following
the instructions under
“Validating the Analyzer”
in the Firmware
Manual.
Disposal of Used Scrubbers
1.
Discard used scrubber and scrubber indicator in appropriate
receptacle.
Depleted H
2
S scrubbers and scrubber indicators contain
predominantly Copper (II) Sulfide [CAS# 1317-40-4] with some
remaining Copper (II) Oxide [CAS# 1317-38-0] and basic cupric
carbonate [CAS# 12069-69-1], each of which are odorless dark
powders that require few special precautions other than avoiding
contact with the internal substances, keeping the scrubber tightly
sealed and protecting the contents against humidity.
Содержание SpectraSensors SS2100r
Страница 1: ...SS2100r H2S Analyzer Rack Mount Hardware Manual P N 4900002189 rev D...
Страница 2: ......
Страница 4: ......
Страница 8: ...SS2100r H2S Analyzer iv 4900002189 rev D 3 7 18 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 10: ...SS2100r H2S Analyzer vi 4900002189 rev D 3 7 18 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 18: ...SS2100r H2S Analyzer 1 8 4900002189 rev D 3 7 18 Figure 1 5 View of front and rear panels of the analyzer...
Страница 19: ...Introduction Hardware Manual 1 9 Figure 1 6 View of inside the analyzer enclosure with top panels removed...
Страница 25: ...Introduction Hardware Manual 1 15 Figure 1 10 Label placement on the inside of the analyzer enclosure...
Страница 34: ...SS2100r H2S Analyzer 2 8 4900002189 rev D 3 7 18 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 44: ...SS2100r H2S Analyzer A 4 4900002189 rev D 3 7 18 THIS PAGE INTENTIONALLY LEFT BLANK...