SS2100r H
2
S Analyzer
2
–
6
4900002189 rev. D 3-7-18
4.
For 4-20 mA sink, carefully replace the jumper to connect the center
pin with point “P.”
5.
Reconnect power to the analyzer. Confirm the 4 mA (min.) and
20 mA (max.) points (refer to
"Calibrating the analog output"
below).
6.
Replace and the left side enclosure cover.
Calibrating the analog output
1.
Connect a calibrator and digital multi-meter into the circuit.
2.
On the analyzer keypad, press
#2
(
Mode 2
), the password (3142)
and
*
.
3.
Continue pressing the
*
key until the
4-20 mA Test
parameter
displays.
Needle nose pliers may be required to remove the jumper
.
Figure 2–2
Analyzer 4-20 mA board
JMP1
POTENTIOMETERS
<SET PARAMETER MODE>
AO 4-20 mA Test
0.00000
Enter a value (%)
Содержание SpectraSensors SS2100r
Страница 1: ...SS2100r H2S Analyzer Rack Mount Hardware Manual P N 4900002189 rev D...
Страница 2: ......
Страница 4: ......
Страница 8: ...SS2100r H2S Analyzer iv 4900002189 rev D 3 7 18 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 10: ...SS2100r H2S Analyzer vi 4900002189 rev D 3 7 18 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 18: ...SS2100r H2S Analyzer 1 8 4900002189 rev D 3 7 18 Figure 1 5 View of front and rear panels of the analyzer...
Страница 19: ...Introduction Hardware Manual 1 9 Figure 1 6 View of inside the analyzer enclosure with top panels removed...
Страница 25: ...Introduction Hardware Manual 1 15 Figure 1 10 Label placement on the inside of the analyzer enclosure...
Страница 34: ...SS2100r H2S Analyzer 2 8 4900002189 rev D 3 7 18 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 44: ...SS2100r H2S Analyzer A 4 4900002189 rev D 3 7 18 THIS PAGE INTENTIONALLY LEFT BLANK...