Introduction
Hardware Manual
1
–
3
controlled by microprocessor-based electronics with embedded software that
incorporates advanced operational and data processing algorithms.
Sample Conditioning System
Sample conditioning is included in the system and has been specifically
designed to deliver an optimum sample stream that is representative of the
process systems stream at the time of sampling. Most SS2100r analyzer
systems are configured for use at extractive natural gas sampling stations.
Determining Firmware Version
When the analyzer is powered on for the first time, the firmware version will
display on the system LCD for approximately seven seconds. Refer to
“Powering Up the Analyzer”
in the Firmware Manual for this analyzer for
operational instructions. The firmware version for each analyzer is also listed
on the analyzer calibration certificate.
How the Analyzers Work
The SS2100r analyzers employ tunable diode laser absorption spectroscopy
(TDLAS) to detect the presence of trace substances in process gases.
Absorption spectroscopy is a widely used technique for sensitive trace species
detection. Because the measurement is made in the volume of the gas, the
response is much faster, more accurate and significantly more reliable than
traditional surface-based sensors that are subject to surface contamination.
In its simplest form, a diode laser absorption spectrometer typically consists of
a sample cell with a mirror at one end, and a mirror or window at the opposite
end, through which the laser beam can pass. Refer to Figure 1–1. The laser
beam enters the cell and reflects off the mirror(s) making one or more trips
through the sample gas and eventually exiting the cell where the remaining
beam intensity is measured by a detector. With the SS2100r analyzers, sample
gas flows continuously through the sample cell ensuring that the sample is
always representative of the flow in the main pipe.
Due to their inherent structure, the molecules in the sample gas each have
characteristic natural frequencies (or resonances). When the output of the
laser is tuned to one of those natural frequencies, the molecules with that
particular resonance will absorb energy from the incident beam. That is, as the
beam of incident intensity,
I
0
(
)
,
passes through the sample, attenuation occurs
via absorption by the trace gas with absorption cross section
(
). According
to the Beer-Lambert absorption law, the intensity remaining,
I
(
), as measured
by the detector at the end of the beam path of length / (cell length x number
of passes), is given by
,
(1)
where
N
represents the species concentration. Thus, the ration of the
absorption measured when the laser is tuned on-resonance versus off-
I
I
0
exp
lN
–
=
Содержание SpectraSensors SS2100r
Страница 1: ...SS2100r H2S Analyzer Rack Mount Hardware Manual P N 4900002189 rev D...
Страница 2: ......
Страница 4: ......
Страница 8: ...SS2100r H2S Analyzer iv 4900002189 rev D 3 7 18 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 10: ...SS2100r H2S Analyzer vi 4900002189 rev D 3 7 18 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 18: ...SS2100r H2S Analyzer 1 8 4900002189 rev D 3 7 18 Figure 1 5 View of front and rear panels of the analyzer...
Страница 19: ...Introduction Hardware Manual 1 9 Figure 1 6 View of inside the analyzer enclosure with top panels removed...
Страница 25: ...Introduction Hardware Manual 1 15 Figure 1 10 Label placement on the inside of the analyzer enclosure...
Страница 34: ...SS2100r H2S Analyzer 2 8 4900002189 rev D 3 7 18 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 44: ...SS2100r H2S Analyzer A 4 4900002189 rev D 3 7 18 THIS PAGE INTENTIONALLY LEFT BLANK...