Solitrend MMP60

Basic safety instructions

Hauser

5

2.2  

Intended use

Application and media

The device described in this manual is intended for the continuous moisture measurement of

a wide variety of materials. Because of its operating frequency of approx. 1 GHz the device can

also be used outside of closed metal vessels.
If operated outside of closed vessels, the device must be mounted in accordance with the

instructions in the 

Mounting

 section. The operation of the devices does not present any

health risk. If the limit values specified in the 

Technical data

 and the conditions listed in the

instructions and additional documentation are observed, the measuring device may be used

for the following measurements only:
• Measured process variables: material moisture, material conductivity and material

temperature
To ensure that the device remains in proper condition throughout the operating period:

Use the device only for media to which the process-wetted materials are sufficiently

resistant.

Observe the limit values in the "Technical data".

Incorrect use

The manufacturer is not liable for damage caused by improper or non-intended use.
Clarification of borderline cases:

With regard to special fluids and media used for cleaning, the manufacturer is glad to

assist in clarifying the corrosion resistance of materials in contact with the fluid, but does

not accept any warranty or liability.

Residual risks

Due to the transfer of heat from the process and power dissipation within the electronics, the

temperature of the electronics housing and the assemblies contained therein may increase to

up to 70 °C (158 °F) during operation. The device may reach a temperature close to the

medium temperature during operation.
Danger of burns from contact with surfaces!

In the event of high medium temperatures, ensure protection against contact to prevent

burns.

2.3  

Workplace safety

For work on and with the device:

Wear the required personal protective equipment according to federal/national

regulations.

2.4  

Operational safety

Risk of injury!

Operate the device only if it is in proper technical condition, free from errors and faults.

The operator is responsible for interference-free operation of the device.

Содержание Solitrend MMP60

Страница 1: ...tructions they are not a substitute for the Operating Instructions pertaining to the device For detailed information refer to the Operating Instructions and other documentation Available for all devic...

Страница 2: ...Solitrend MMP60 2 Endress Hauser Order code Ext ord cd Ser no www endress com deviceviewer Endress Hauser Operations App XXXXXXXXXXXX XXXXX XXXXXX XXX XXXX XX Serial number 1 3 2 A0023555...

Страница 3: ...7 4 2 Product identification 7 4 3 Manufacturer s address 8 4 4 Storage transport 8 5 Mounting 8 5 1 Mounting requirements 8 5 2 Mounting the device 8 5 3 Post mounting check 9 6 Electrical connection...

Страница 4: ...ollowing requirements Trained qualified specialists must have a relevant qualification for this specific function and task Personnel must be authorized by the plant owner operator Be familiar with fed...

Страница 5: ...t Observe the limit values in the Technical data Incorrect use The manufacturer is not liable for damage caused by improper or non intended use Clarification of borderline cases With regard to special...

Страница 6: ...ements has been tested and left the factory in a condition in which it is safe to operate It meets the general safety standards and legal requirements It also complies with the EU directives listed in...

Страница 7: ...ditions is not met please contact the manufacturer s sales office 4 2 Product identification The device can be identified in the following ways Nameplate specifications Extended order code with breakd...

Страница 8: ...lace of installation to ensure that the material flow and therefore the bulk density over the measurement surface is constant The measuring field of the device must be completely covered by material a...

Страница 9: ...ns Unit of measurement mm in 5 2 1 Mounting the housing with the remote electronics module The housing with the remote electronics module can be mounted with two screws M5 163 6 42 52 2 05 8 0 3 1 A00...

Страница 10: ...ured 6 Electrical connection 6 1 Connecting requirements 6 1 1 Cable specification Connecting cables with a pre assembled 10 pin socket on the device side are available in different standard lengths 4...

Страница 11: ...WH G RS485 B must be enabled Wire color brown BN C IMP Bus RT Wire color gray GY pink PK see figure below J IMP Bus COM Wire color blue BU red RD see figure below K 2nd analog positive Wire color pin...

Страница 12: ...g output The device can be set to 0 to 20 mA or 4 to 20 mA The current output can also be set inversely to 20 to 0 mA or 20 to 4 mA for special controllers and applications The analog outputs can be s...

Страница 13: ...10 0 to 20 or 0 to 30 8 1 1 Possible settings There are several possible settings for the analog outputs Analog outputs Options 0 to 20 mA 4 to 20 mA The current output can also be set inversely for...

Страница 14: ...setting is saved to the non volatile memory of the device 8 2 1 Operating mode The device is supplied from the factory with the CH mode for applications in the construction industry and with the CA mo...

Страница 15: ...o discharge hatch The CH mode performs filtering automatically This allows drip water that forms in the silo to be filtered out of the measured value for example 8 3 Calibration curve set A for genera...

Страница 16: ...6 Calibration curve set A Cal 1 Cal 2 Cal 3 Cal 4 Cal 5 Cal 6 Cal 7 Cal 10 Cal 11 H Gravimetric moisture t Radar transit time picoseconds 1 Cal 1 Universal sand gravel grit 2 Cal 2 Sand 1 6 3 Cal 3 S...

Страница 17: ...urves Cal 1 to Cal 15 for various materials that are saved and can be selected in the device The gravimetric moisture H is indicated as a percentage on the y axis and the associated radar transit time...

Страница 18: ...el shift offset in the PLC The parameter has different names depending on the PLC e g initial load zero point offset measuring range etc Perform a parallel shift offset in the PLC Contact the PLC manu...

Страница 19: ...se grained or hydrophobic materials water can escape directly onto the measurement surface and therefore cause a high moisture value Enter limit values in the PLC Contact the PLC manufacturer Incorrec...

Страница 20: ...www addresses endress com 71550111 71550111...

Отзывы: