Solitrend MMP60

Commissioning

Hauser

13

Moisture, temperature/conductivity

• Output 1: moisture in % (variable setting)

• Output 2: material temperature 0 to 100 °C (32 to 212 °F) and conductivity 0 to 20 mS/cm

with automatic window switching.

It is also possible to divide output 2 into two ranges to output both the conductivity and the

temperature, namely the 4 to 11 mA range for the temperature and the 12 to 20 mA range

for the conductivity. Output 2 automatically switches between these two windows every 5 s.

Output 1 can also be scaled at the factory or subsequently scaled as required (variably)

using the remote display (optionally available), e.g. 0 to 10 %, 0 to 20 % or 0 to 30 %

8.1.1  

Possible settings

There are several possible settings for the analog outputs:

Analog outputs
Options:

• 0 to 20 mA

• 4 to 20 mA

The current output can also be set inversely for special controllers and applications.
• 20 to 0 mA

• 20 to 4 mA

Channels of the analog outputs

The analog outputs can be set differently to the following possible options:

Moisture, temperature

Output 1 for moisture, output 2 for material temperature.

Moisture, conductivity

Output 1 for moisture, output 2 for conductivity in the range from 0 to 20 mS/cm

(factory setting)

Moisture, temperature/conductivity

Output 1 for moisture, output 2 for material temperature and conductivity with

automatic window switching.

Moisture range

The moisture range and the temperature range at outputs 1 and 2 can be individually

configured.

The moisture range may not exceed 100 %.

Содержание Solitrend MMP60

Страница 1: ...tructions they are not a substitute for the Operating Instructions pertaining to the device For detailed information refer to the Operating Instructions and other documentation Available for all devic...

Страница 2: ...Solitrend MMP60 2 Endress Hauser Order code Ext ord cd Ser no www endress com deviceviewer Endress Hauser Operations App XXXXXXXXXXXX XXXXX XXXXXX XXX XXXX XX Serial number 1 3 2 A0023555...

Страница 3: ...7 4 2 Product identification 7 4 3 Manufacturer s address 8 4 4 Storage transport 8 5 Mounting 8 5 1 Mounting requirements 8 5 2 Mounting the device 8 5 3 Post mounting check 9 6 Electrical connection...

Страница 4: ...ollowing requirements Trained qualified specialists must have a relevant qualification for this specific function and task Personnel must be authorized by the plant owner operator Be familiar with fed...

Страница 5: ...t Observe the limit values in the Technical data Incorrect use The manufacturer is not liable for damage caused by improper or non intended use Clarification of borderline cases With regard to special...

Страница 6: ...ements has been tested and left the factory in a condition in which it is safe to operate It meets the general safety standards and legal requirements It also complies with the EU directives listed in...

Страница 7: ...ditions is not met please contact the manufacturer s sales office 4 2 Product identification The device can be identified in the following ways Nameplate specifications Extended order code with breakd...

Страница 8: ...lace of installation to ensure that the material flow and therefore the bulk density over the measurement surface is constant The measuring field of the device must be completely covered by material a...

Страница 9: ...ns Unit of measurement mm in 5 2 1 Mounting the housing with the remote electronics module The housing with the remote electronics module can be mounted with two screws M5 163 6 42 52 2 05 8 0 3 1 A00...

Страница 10: ...ured 6 Electrical connection 6 1 Connecting requirements 6 1 1 Cable specification Connecting cables with a pre assembled 10 pin socket on the device side are available in different standard lengths 4...

Страница 11: ...WH G RS485 B must be enabled Wire color brown BN C IMP Bus RT Wire color gray GY pink PK see figure below J IMP Bus COM Wire color blue BU red RD see figure below K 2nd analog positive Wire color pin...

Страница 12: ...g output The device can be set to 0 to 20 mA or 4 to 20 mA The current output can also be set inversely to 20 to 0 mA or 20 to 4 mA for special controllers and applications The analog outputs can be s...

Страница 13: ...10 0 to 20 or 0 to 30 8 1 1 Possible settings There are several possible settings for the analog outputs Analog outputs Options 0 to 20 mA 4 to 20 mA The current output can also be set inversely for...

Страница 14: ...setting is saved to the non volatile memory of the device 8 2 1 Operating mode The device is supplied from the factory with the CH mode for applications in the construction industry and with the CA mo...

Страница 15: ...o discharge hatch The CH mode performs filtering automatically This allows drip water that forms in the silo to be filtered out of the measured value for example 8 3 Calibration curve set A for genera...

Страница 16: ...6 Calibration curve set A Cal 1 Cal 2 Cal 3 Cal 4 Cal 5 Cal 6 Cal 7 Cal 10 Cal 11 H Gravimetric moisture t Radar transit time picoseconds 1 Cal 1 Universal sand gravel grit 2 Cal 2 Sand 1 6 3 Cal 3 S...

Страница 17: ...urves Cal 1 to Cal 15 for various materials that are saved and can be selected in the device The gravimetric moisture H is indicated as a percentage on the y axis and the associated radar transit time...

Страница 18: ...el shift offset in the PLC The parameter has different names depending on the PLC e g initial load zero point offset measuring range etc Perform a parallel shift offset in the PLC Contact the PLC manu...

Страница 19: ...se grained or hydrophobic materials water can escape directly onto the measurement surface and therefore cause a high moisture value Enter limit values in the PLC Contact the PLC manufacturer Incorrec...

Страница 20: ...www addresses endress com 71550111 71550111...

Отзывы: